EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF used with an Olympus video system center, light source...
Olympus Corporation of the Americas began a recall of EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF used with an Olympus video system center, light source, documentation equipment monitors, Endo Therapy... on . FDA classified it as Class II and gives the reason as "New reprocessing instructions for the Olympus TJF-160/VF duodenoscopes, consisting of revised manual cleaning and high level disinfection procedures they plan to implement as well as a new cleaning brush (MAJ-1534)...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1845-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF used with an Olympus video system center, light source, documentation equipment monitors, Endo Therapy accessories and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
- Product code
- FDT Duodenoscope And Accessories, Flexible/Rigid
- Reason for recall
- New reprocessing instructions for the Olympus TJF-160/VF duodenoscopes, consisting of revised manual cleaning and high level disinfection procedures they plan to implement as well as a new cleaning brush (MAJ-1534), which is enclosed with the notification.
- FDA root cause
- Under Investigation by firm
- Action
- Olympus notified customers on 03/14/2016 via an Urgent Safety Notification letter that referenced discontinuation of previously distributed copies of the TJF-160F/VF Reprocessing Manual. Customers were asked to inspect their inventory and discard any existing inventory of TJF-160F/VF Reprocessing Manuals. They were asked to update their information with the enclosed TJF-160F/VF Reprocessing Manual, which contains new brushing and flushing steps for the TJF-160F/VF elevator mechanisms and the forceps elevator recesses. Instuctions for cleaning and a new cleaning brush (used for cleaning the device) was included with the notification. Customers were asked to ensure all reprocessing personnel are completely knowledgeable and throughly training on the new reprocessing instructions in the new Reprocessing Manual.
- Distribution
- Nationwide:
- Product codes and lots
- Models: TJF-160F and TJF-160VF, all serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1845-2016 | Class II | Terminated | Voluntary: Firm initiated |