Z-1845-2016Class II

EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF used with an Olympus video system center, light source...

OLYMPUS Corporation of the Americas / initiated 2016-03-15 / Terminated

Olympus Corporation of the Americas began a recall of EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF used with an Olympus video system center, light source, documentation equipment monitors, Endo Therapy... on . FDA classified it as Class II and gives the reason as "New reprocessing instructions for the Olympus TJF-160/VF duodenoscopes, consisting of revised manual cleaning and high level disinfection procedures they plan to implement as well as a new cleaning brush (MAJ-1534)...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1845-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
OLYMPUS Corporation of the Americas
Product
EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF used with an Olympus video system center, light source, documentation equipment monitors, Endo Therapy accessories and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
Product code
FDT Duodenoscope And Accessories, Flexible/Rigid
Reason for recall
New reprocessing instructions for the Olympus TJF-160/VF duodenoscopes, consisting of revised manual cleaning and high level disinfection procedures they plan to implement as well as a new cleaning brush (MAJ-1534), which is enclosed with the notification.
FDA root cause
Under Investigation by firm
Action
Olympus notified customers on 03/14/2016 via an Urgent Safety Notification letter that referenced discontinuation of previously distributed copies of the TJF-160F/VF Reprocessing Manual. Customers were asked to inspect their inventory and discard any existing inventory of TJF-160F/VF Reprocessing Manuals. They were asked to update their information with the enclosed TJF-160F/VF Reprocessing Manual, which contains new brushing and flushing steps for the TJF-160F/VF elevator mechanisms and the forceps elevator recesses. Instuctions for cleaning and a new cleaning brush (used for cleaning the device) was included with the notification. Customers were asked to ensure all reprocessing personnel are completely knowledgeable and throughly training on the new reprocessing instructions in the new Reprocessing Manual.
Distribution
Nationwide:
Product codes and lots
Models: TJF-160F and TJF-160VF, all serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1845-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1845-2016Class IITerminatedVoluntary: Firm initiated