Z-1376-2016Class II

da Vinci S Surgical System (IS1200/IS2000/IS3000) user manual, instrument & accessory

Intuitive Surgical, Inc. / initiated 2016-03-16 / Terminated / root cause: design

Intuitive Surgical, Inc. began a recall of da Vinci S Surgical System (IS1200/IS2000/IS3000) user manual, instrument & accessory. Product Usage: The Intuitive Surgical Endoscopic Instrument Control... on . FDA classified it as Class II and gives the reason as "Intuitive Surgical has found that the use of instruments other than the Cardiac Probe Grasper (e.g. Large Needle Driver Instrument) to manipulate the ablation probes during cardiac surgery can result in an increase of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1376-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci S Surgical System (IS1200/IS2000/IS3000) user manual, instrument & accessory. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecological laparoscopic surgical procedures, general noncardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization.It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Intuitive Surgical has found that the use of instruments other than the Cardiac Probe Grasper (e.g. Large Needle Driver Instrument) to manipulate the ablation probes during cardiac surgery can result in an increase of microscopic metallic particulates generated from the contact with the probes.
Root cause tag
design
FDA root cause
Device Design
Action
Intuitive sent a Field Safety Notice/Urgent Medical Device Correction letters dated March 22, 2016 by trackable mail method to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to 1. In addition to existing instructions for use, observe the following warning when manipulating the cardiac ablation probe, Use only the Cardiac Probe Grasper to manipulate a cardiac ablation probe. 2. Distribute a copy of this letter to all da Vinci S, Si, and Xi users at your facility. 3. Place a copy of this letter with your user manual. 4. Complete the attached Acknowledgement Form and return it to Intuitive Surgical as instructed. 5. Retain a copy of this letter and the Acknowledgement Form for your files. Intuitive Surgical will provide a user manual addendum to incorporate the warning related to this risk. For questions contact your Clinical Sales Representative or contact Intuitive Surgical Customer Service.
Quantity in commerce
248 - all models
Distribution
Worldwide Distribution - US Nationwide in the countries Australia, Belgium, Brazil, Canada, Chile, Finland, France, Germany, Greece, Iceland, India, Israel, Italy, Netherlands, Qatar, Russia, Singapore, South Korea, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Product codes and lots
Model 550675; IS1200/IS2000/IS3000 - User Manual, Instrument & Accessory - all languages.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1376-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1376-2016Class IITerminatedVoluntary: Firm initiated
Z-1377-2016Class IITerminatedVoluntary: Firm initiated