Z-1576-2016Class II

Drew Scientific Inc. Drew3-PAC. This D3 Reagent Pack includes: diluent reagent 5L, enzymatic cleaner 750 mL...

JAS Diagnostics/Drew Scientific / initiated 2016-03-10 / Terminated / root cause: manufacturing process

Jas Diagnostics/Drew Scientific began a recall of Drew Scientific Inc. Drew3-PAC. This D3 Reagent Pack includes: diluent reagent 5L, enzymatic cleaner 750 mL, and a lysing reagent 125 mL on . FDA classified it as Class II and gives the reason as "The reagent pack may result in elevated platelet background during Drew3 Hematology Analyzer routine check". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1576-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
JAS Diagnostics/Drew Scientific
Product
Drew Scientific Inc. Drew3-PAC. This D3 Reagent Pack includes: diluent reagent 5L, enzymatic cleaner 750 mL, and a lysing reagent 125 mL.
Product code
GKZ Counter, Differential Cell
Reason for recall
The reagent pack may result in elevated platelet background during Drew3 Hematology Analyzer routine check.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
JAS Diagnostics/ Drew Scientifics sent an Important Product Advisory letter dated March 10, 2016, to all affected customers. The domestic customers were notified via letter and international customers were notified via email. Customers were instructed to contact Drew Scientific Inc. Customer Service or Technical Service at 1-800-433-0945 for product replacement. The firm suggested a temporary workaround of performing five background runs three times per day in order to ensure background counts are running within specification.
Quantity in commerce
7,691
Distribution
Worldwide Distribution - US Distribution including the states of : AL, AR, CA, FL, GA, IA, IN, KS, KY, LA, MA, MN, MO, MS, NC, NE, NY, SC, TN, and TX. and to the countries of : Bolivia, Cairo, Columbia, Ecuador, Italy, Moscow, Netherlands, Pakistan, Peru, Slovenia, Spain, and Venezuela.
Product codes and lots
Lot # 0825, exp 01/06/2018; Lot # 0826, exp 01/06/2018; Lot # 0827, exp 01/06/2018; Lot # 0828, exp 01/21/2018; Lot # 0829, exp 01/21/2018; Lot # 0830, exp 01/21/2018; Lot # 0831, exp 01/21/2018; and Lot # 0832, exp 01/21/2018.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1576-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1576-2016Class IITerminatedVoluntary: Firm initiated