Drew Scientific Inc. Drew3-PAC. This D3 Reagent Pack includes: diluent reagent 5L, enzymatic cleaner 750 mL...
Jas Diagnostics/Drew Scientific began a recall of Drew Scientific Inc. Drew3-PAC. This D3 Reagent Pack includes: diluent reagent 5L, enzymatic cleaner 750 mL, and a lysing reagent 125 mL on . FDA classified it as Class II and gives the reason as "The reagent pack may result in elevated platelet background during Drew3 Hematology Analyzer routine check". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1576-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- JAS Diagnostics/Drew Scientific
- Product
- Drew Scientific Inc. Drew3-PAC. This D3 Reagent Pack includes: diluent reagent 5L, enzymatic cleaner 750 mL, and a lysing reagent 125 mL.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- The reagent pack may result in elevated platelet background during Drew3 Hematology Analyzer routine check.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- JAS Diagnostics/ Drew Scientifics sent an Important Product Advisory letter dated March 10, 2016, to all affected customers. The domestic customers were notified via letter and international customers were notified via email. Customers were instructed to contact Drew Scientific Inc. Customer Service or Technical Service at 1-800-433-0945 for product replacement. The firm suggested a temporary workaround of performing five background runs three times per day in order to ensure background counts are running within specification.
- Quantity in commerce
- 7,691
- Distribution
- Worldwide Distribution - US Distribution including the states of : AL, AR, CA, FL, GA, IA, IN, KS, KY, LA, MA, MN, MO, MS, NC, NE, NY, SC, TN, and TX. and to the countries of : Bolivia, Cairo, Columbia, Ecuador, Italy, Moscow, Netherlands, Pakistan, Peru, Slovenia, Spain, and Venezuela.
- Product codes and lots
- Lot # 0825, exp 01/06/2018; Lot # 0826, exp 01/06/2018; Lot # 0827, exp 01/06/2018; Lot # 0828, exp 01/21/2018; Lot # 0829, exp 01/21/2018; Lot # 0830, exp 01/21/2018; Lot # 0831, exp 01/21/2018; and Lot # 0832, exp 01/21/2018.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1576-2016 | Class II | Terminated | Voluntary: Firm initiated |