RTI Surgical Streamline Occipito-Cervico-Thoracic (OCT) Final Driver, Common Name: Screw Driver
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) began a recall of RTI Surgical Streamline Occipito-Cervico-Thoracic (OCT) Final Driver, Common Name: Screw Driver on . FDA classified it as Class II and gives the reason as "Multiple lots of the 26-FNLDRIVER, may have reduced mechanical strength, resulting in driver tip fracturing during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1393-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
- Product
- RTI Surgical Streamline Occipito-Cervico-Thoracic (OCT) Final Driver, Common Name: Screw Driver
- Product code
- KWP Appliance, Fixation, Spinal Interlaminal
- Reason for recall
- Multiple lots of the 26-FNLDRIVER, may have reduced mechanical strength, resulting in driver tip fracturing during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- RTI Surgical Inc. initiated a voluntary recall of RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Final Driver, via written notices sent by e-mail and FedEx on 03/18/2016. Customers are requested to review all inventory, quarantine products listed in the recall, and to immediately return recalled products with the recall reply form, and the shipping labels and authorization numbers provided within 5 days of receiving notice. Replacement drivers are being sent to customers. Customers may contact Dan Nelson, Manager of Quality Systems, Phone: 906-226-4489, Fax: 906-226-4455, e-mail: [email protected].
- Distribution
- Domestic: AL, CO, CT, FL, IN, ID, IL, MI, MS, MO, NC, NY, TX Foreign: None VA/DOD: None
- Product codes and lots
- 26-FNLDRIVER, Lot Numbers: 177091, 177093, 183817, 207727, 223926, 227291, 229134, 230523, 232262, 233074, 234205
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1393-2016 | Class II | Terminated | Voluntary: Firm initiated |