Boston Scientific Ultraflex Tracheobronchial Distal Release Stent System 12mm X 30mm Material Number...
Boston Scientific Corporation began a recall of Boston Scientific Ultraflex Tracheobronchial Distal Release Stent System 12mm X 30mm Material Number: M00576550 The stent system is a permanently implanted... on . FDA classified it as Class II and gives the reason as "Mislabeled: 10mm diameter x 30mm length Ultraflex Tracheobronchial stents were labeled as 12mm diameter x 30mm". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1567-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Ultraflex Tracheobronchial Distal Release Stent System 12mm X 30mm Material Number: M00576550 The stent system is a permanently implanted expandable metal stent designed for palliation of tracheobronchial (TB) strictures. The system consists of a flexible delivery catheter preloaded with an expandable metal stent. The stent is offered in a variety of sizes of different diameters and lengths
- Product code
- JCT Prosthesis, Tracheal, Expandable
- Reason for recall
- Mislabeled: 10mm diameter x 30mm length Ultraflex Tracheobronchial stents were labeled as 12mm diameter x 30mm
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Boston Scientific issued notification letters dated March 17, 2016. The notification instructs the customer to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form. Customers with questions should call 866-868-4004. For questions regarding this recall call 508-382-9555.
- Quantity in commerce
- 4 units
- Distribution
- Nationwide Distribution to AL, CA, MA, and TX.
- Product codes and lots
- Lot Number: 18876596 Exp Date: February 1, 201 8
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1567-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY