Poly-bags containing the strap labeled with "Adult" sticker. Each strap kit contains adjustable length leg...

Ferno-Washington Inc / initiated 2016-03-10 / Terminated / root cause: manufacturing process

Ferno-Washington Inc began a recall of Poly-bags containing the strap labeled with "Adult" sticker. Each strap kit contains adjustable length leg straps, ankle wrap, a thigh pad and a heel stand... on . FDA classified it as Class II and gives the reason as "On March 9, 2016, the firm was contacted regarding the possibility of incorrect labeling on the outside poly- bag of the traction splint straps. The firm investigated and determined its supplier had labeled the outside...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1604-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ferno-Washington Inc
Product
Poly-bags containing the strap labeled with "Adult" sticker. Each strap kit contains adjustable length leg straps, ankle wrap, a thigh pad and a heel stand. The strap kits are applied to the traction splint frame to immobilize and provide traction for a patients leg.
Product code
NZD Stretcher, Patient Restraint
Reason for recall
On March 9, 2016, the firm was contacted regarding the possibility of incorrect labeling on the outside poly- bag of the traction splint straps. The firm investigated and determined its supplier had labeled the outside packaging for the pediatric and adult traction splints straps incorrectly. The packaging labeled Adult contained a pediatric splint strap and the packaging labeled Pediatric conta
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Ferno-Washington, Inc , has contacted all customers via telephone on March 10, 2016. All products will be replaced with new products. For further questions, please call: (937) 283-2885.
Quantity in commerce
270
Distribution
Worldwide Distribution - US Distribution to the states of : AK ,AZ ,CA, CO, FL ,GA ,HI, IL ,IN ,KY ,MA ,MI ,MS ,NC ,NE ,NJ ,NY ,OH,ON,TX ,VA ,WI , and WV. There are no government accounts for this recall. There is one Canadian account for this recall. There are no Mexican accounts for this recall. and to the countries of : UAE, MALAYSIA, UNITED KINGDOM, SLOVAKIA, SOUTH AFRICA, QATAR and AUSTRALIA..
Product codes and lots
Strap Kit Adult Model 444