ADVIA CHEMISTRY TRIGLYCERIDES_2 (TRIG_2) REAGENT & ADVIA CHEMISTRY TRIGLYCERIDES_2 Concentrated (TRIG_c)...
Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA CHEMISTRY TRIGLYCERIDES_2 (TRIG_2) REAGENT & ADVIA CHEMISTRY TRIGLYCERIDES_2 Concentrated (TRIG_c) REAGENT on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1657-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- ADVIA CHEMISTRY TRIGLYCERIDES_2 (TRIG_2) REAGENT & ADVIA CHEMISTRY TRIGLYCERIDES_2 Concentrated (TRIG_c) REAGENT
- Product code
- CDT Lipase Hydrolysis/Glycerol Kinase Enzyme, Triglycerides
- Reason for recall
- Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens mailed an Urgent Medical Device Correction letter to customers making them aware of the problem. Special instructions were also given for the customer to review the letter with their Medical Director and ensure Venipuncture occurs before drug administration of NAC or Metamizole as indicated in the letter. Baseline assay values before administration of NAC or Metamizole therapy would not be affected. Customers were asked to complete and return the Field Correction Effectiveness Check Form attached with the letter within 30 days. Furthermore, if customers received any complaints of illness or adverse events associated with the specified lots, they were to immediately contact their local Siemens Customer Care Center or local Siemens technical support representative.
- Quantity in commerce
- 137422 units total
- Distribution
- Nationwide: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, St. Thomas, VI, Guam, Pacific Mariana Islands
- Product codes and lots
- Device Listing: D126071, Triglyceride_2, Catalog # 74082 Device Listing: D211156, Triglyceride Concentrated, Catalog # 74084
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1651-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1652-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1653-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1654-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1655-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1656-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1657-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1658-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1659-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1660-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1661-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1662-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1663-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1664-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1665-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1666-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
