Z-1659-2016Class II

Dimension Assays: Direct HDL Cholesterol (AHDL)

Siemens Healthcare Diagnostics Inc. / initiated 2016-03-17 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Assays: Direct HDL Cholesterol (AHDL) on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1659-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Assays: Direct HDL Cholesterol (AHDL)
Product code
JHM Colorimetric Method, Lipoproteins
Reason for recall
Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific
Root cause tag
design
FDA root cause
Device Design
Action
Siemens mailed an Urgent Medical Device Correction letter to customers making them aware of the problem. Special instructions were also given for the customer to review the letter with their Medical Director and ensure Venipuncture occurs before drug administration of NAC or Metamizole as indicated in the letter. Baseline assay values before administration of NAC or Metamizole therapy would not be affected. Customers were asked to complete and return the Field Correction Effectiveness Check Form attached with the letter within 30 days. Furthermore, if customers received any complaints of illness or adverse events associated with the specified lots, they were to immediately contact their local Siemens Customer Care Center or local Siemens technical support representative.
Quantity in commerce
137422 units total
Distribution
Nationwide: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, St. Thomas, VI, Guam, Pacific Mariana Islands
Product codes and lots
Device Listing: D 044702 - Direct HDL Cholesterol (AHDL), Catalog # DF48B

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1659-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1651-2016Class IITerminatedVoluntary: Firm initiated
Z-1652-2016Class IITerminatedVoluntary: Firm initiated
Z-1653-2016Class IITerminatedVoluntary: Firm initiated
Z-1654-2016Class IITerminatedVoluntary: Firm initiated
Z-1655-2016Class IITerminatedVoluntary: Firm initiated
Z-1656-2016Class IITerminatedVoluntary: Firm initiated
Z-1657-2016Class IITerminatedVoluntary: Firm initiated
Z-1658-2016Class IITerminatedVoluntary: Firm initiated
Z-1659-2016Class IITerminatedVoluntary: Firm initiated
Z-1660-2016Class IITerminatedVoluntary: Firm initiated
Z-1661-2016Class IITerminatedVoluntary: Firm initiated
Z-1662-2016Class IITerminatedVoluntary: Firm initiated
Z-1663-2016Class IITerminatedVoluntary: Firm initiated
Z-1664-2016Class IITerminatedVoluntary: Firm initiated
Z-1665-2016Class IITerminatedVoluntary: Firm initiated
Z-1666-2016Class IITerminatedVoluntary: Firm initiated