Z-1283-2016Class II

AXIOM Artis zee/zeego, system, x-ray, angiographic

Siemens Medical Solutions USA, Inc. / initiated 2016-02-18 / Terminated

Siemens Medical Solutions USA, Inc began a recall of AXIOM Artis zee/zeego, system, x-ray, angiographic Product Usage: AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single... on . FDA classified it as Class II and gives the reason as "Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can result in a cooling system failure and subsequent system failure. This leak can occur sporadically in...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1283-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
AXIOM Artis zee/zeego, system, x-ray, angiographic Product Usage: AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can result in a cooling system failure and subsequent system failure. This leak can occur sporadically in affected systems.
FDA root cause
Unknown/Undetermined by firm
Action
Siemens sent an Important Customer Safety Notice to customers on February 18, 2016 informing them that their service organization will contact them to arrange a date to perform the corrective action (replacement of affected plastic couplings with metal couplings to prevent fluid leaks). Customers were advised to contact Siemens service organization for an earlier appointment at 1-800-888-7436.
Quantity in commerce
986 systems in total
Distribution
US Nationwide Distribution in the states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY including Puerto Rico.
Product codes and lots
Serial Numbers:  Artis Q biplane Serial #  121061, 121089, 121094, 121006, 121013, 121040, 121062, 121082, 121011, 121039, 121134, 121053, 121054, 121014, 121112, 121051, 121004, 121133, 121018, 121036, 121137, 121012, 121049, 121078, 121019, 121010, 121139, 121081, 121096, 121015, 121115, 121016, 121138, 121020, 121140, 121106, 121120, 121122, 121121, 121109, 121144, 121047, 121088, 121027, 121136, 121141, 121069, 121009, 121057, 121056, 121034, 121008, 121058, 121085, 121071, 121023, 121079, 121041, 121037, 121046, 121072

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1283-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1283-2016Class IITerminatedVoluntary: Firm initiated
Z-1284-2016Class IITerminatedVoluntary: Firm initiated