Z-1503-2016Class II

Ad-Tech Electrode Connection System/Cables. Lightweight TECH-ATTACH Cable Models: 1. L-DCL-4DINX (4 contacts...

AD-Tech Medical Instrument Corp. / initiated 2016-03-29 / Terminated / root cause: packaging

Ad-Tech Medical Instrument Corporation began a recall of Ad-Tech Electrode Connection System/Cables. Lightweight TECH-ATTACH Cable Models: 1. L-DCL-4DINX (4 contacts: 1 terminal block with 4 connectors each); 2... on . FDA classified it as Class II and gives the reason as "Ad-Tech is recalling unopened/unused sterile cables intended to be used only with all compatible Ad-Tech electrodes for the purpose of connecting these electrodes to third party monitors/stimulators. This recall has...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1503-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AD-Tech Medical Instrument Corp.
Product
Ad-Tech Electrode Connection System/Cables. Lightweight TECH-ATTACH Cable Models: 1. L-DCL-4DINX (4 contacts: 1 terminal block with 4 connectors each); 2. L-DCL-6DINX (6 contacts: 1 terminal block with 6 connectors each); 3. L-DCL-8DINX (8 contacts: 1 terminal block with 8 connectors each); 4. L-DCL-16BDINX (16 contacts: 1 terminal block with 16 connectors each); 5. L-DCL-32BDINX (32 contacts: 2 terminal blocks with 16 connectors each); 6. L-DCL-64BDINX (64 contacts: 4 terminal blocks with 16 connectors each). FO TECH-ATTACH Connection Cable model FO-LDC-4DINX (4 contacts: 1 terminal block with 4 connectors each). Tech-Attach: these cables require the use of an additional component - the connector block. The tail of the electrode is inserted into the block and the block in turn mates with the cable. A key-pin is used to ensure proper orientation. The cable terminated in EEG compatible connector sockets. Lightweight CABRIO Cable, model L-SRL-64BDINX (64 contacts: 4 terminal blocks with 16 connectors each). Cabrio cables are similar to Tech-Attach cables except that the block portion is pre-attached. The tail of an electrode is inserted into this portion which then hinges shut.
Product code
IKD Cable, Electrode
Reason for recall
Ad-Tech is recalling unopened/unused sterile cables intended to be used only with all compatible Ad-Tech electrodes for the purpose of connecting these electrodes to third party monitors/stimulators. This recall has been initiated due to package failures (i.e. compromised sterile barriers) that were identified during the execution of a shipping simulation investigation.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Consignees were sent a "Medical Device Recall" letter dated March 29, 2016. The letter described the problem and the product being recalled. Advised consignees to immediately examine their inventory, quarantine and return the product to AD-Tech. If the product has been further distributed identify customers and notify them of the recall. Requested consignees to complete and return the "Medical Device Recall Return Response" form to their Ad-Tech Clinical Specialist at [email protected]; FAX: 262-634-5668 or Telephone 262-634-1555. For questions contact Ad-Tech Clinical Specialist. .
Quantity in commerce
197
Distribution
Distributed in the states of GA and MI, and the countries of FRANCE, UNITED KINGDOM, COLOMBIA, and BRAZIL.
Product codes and lots
Lot Numbers: 208140507, 208140552, 208140556, 208140559, 208140560, 208140561, 208140562, 208140563, 208140564R, 208140579, 208140581, 208140586, 208140588R, 208140590, 208140604.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1503-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1503-2016Class IITerminatedVoluntary: Firm initiated