Z-1577-2016Class I

Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor, Battery Packs Model Number...

Medtronic, Inc. / initiated 2016-04-13 / Terminated

Medtronic began a recall of Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor, Battery Packs Model Number - 016400, 010520. Battery Pack used with the... on . FDA classified it as Class I and gives the reason as "Medtronic Respiratory & Monitoring Solutions Announces a Voluntary Recall of a Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor Due to the Battery Pack Might Experience Thermal Runaway with...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1577-2016
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor, Battery Packs Model Number - 016400, 010520. Battery Pack used with the Capnostream"20 and Capnostream"20p Patient Monitor. The battery pack powers the monitor when AC power is not available or when monitoring is required during intra-hospital transfer of patients. The Capnostream"20 and Capnostream"20p is a combined capnograph/pulse oximeter monitor and its accessories are intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and with continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients in hospitals, hospital-type facilities, and intra-hospital transport environments
Product code
CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Reason for recall
Medtronic Respiratory & Monitoring Solutions Announces a Voluntary Recall of a Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor Due to the Battery Pack Might Experience Thermal Runaway with the Potential of Fire Risk to the Monitor.
FDA root cause
Process change control
Action
Medtronic sent an Urgent Medical Device Recall letter dated April 2016, by postal mail or courier to all affected customers. The letter instructed users to remove and properly dispose of current batteries. If product was further distributed customers were instructed to forward the recall letter. Customers were asked to completed and return the Acknowledgement and Receipt Form by fax or email. Customers with questions should contact Technical Services at 800-635-5267, option 1, ten 1 and 1 or by email at [email protected]. For questions regarding this recall call 303-876-8859.
Quantity in commerce
9,817 batteries (8,170 with monitors and 1,647 as spare parts)
Distribution
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Product was also distributed worldwide. There is 1 VA customer.
Product codes and lots
Product # 016400; All SNs, irrespective of production date. Product # 010520; All SNs with production date starting 14.14

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1577-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1577-2016Class ITerminatedVoluntary: Firm initiated