Z-1370-2016Class II

Boston Scientific, PT Graphix Magnet J-TIP Guidewire with ICE Hydrophilic Coating, Sterile EO, UPN...

Boston Scientific Corporation / initiated 2016-03-28 / Terminated / root cause: manufacturing process

Boston Scientific Corporation began a recall of Boston Scientific, PT Graphix Magnet J-TIP Guidewire with ICE Hydrophilic Coating, Sterile EO, UPN H7491490201MJ2 (Carton Label) ,H7491490201MJo (UPN... on . FDA classified it as Class II and gives the reason as "Boston Scientific is voluntarily implementing a Medical Device Recall of one batch of PT Graphix Magnet J-TIP Guidewires. Boston Scientific has found that, while the batch is labeled as a J-tip model, the guidewires are...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1370-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific, PT Graphix Magnet J-TIP Guidewire with ICE Hydrophilic Coating, Sterile EO, UPN H7491490201MJ2 (Carton Label) ,H7491490201MJo (UPN referenced on the Pouch Label) . Boston Scientific PT Graphix Magnet Guidewire with ICE Hydrophilic Coating is a steerable guidewire available with a nominal diameter of 0.014 in (0.37 mm) and in nominal lengths of 182 cm. PT Graphix Magnet is available with a shapeable Straight Tip or a preformed "J" Tip. The PT Graphix Magnet Guidewire features a 10 or 38 cm polymer sleeve, coated with ICE Hydrophilic Coating, jacketing the distal tapered core wire. The proximal section of the PT Graphix Magnet Guidewire is coated with a fluorinated polymer.
Product code
DQX Wire, Guide, Catheter
Reason for recall
Boston Scientific is voluntarily implementing a Medical Device Recall of one batch of PT Graphix Magnet J-TIP Guidewires. Boston Scientific has found that, while the batch is labeled as a J-tip model, the guidewires are straight and do not have the pre-formed J-tip. The most severe injury expected to occur is a limited procedure delay caused by the need to replace the guidewire with one that has
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Boston Scientific sent an "Urgent Medical Device Recall" letter dated March 28, 2016. The letter was addressed to Cath Lab Materials Manager / Field Action Contact. The letter described the problem and the product affected by the recall. Advised consignees to cease immediately further distribution or use of the product and to notify their customers. They also requested consignees to complete and return "The Reply Verification Tracking Form" For questions contact your local Boston scientific sales Representative of the Boston Scientific Filed Action Center..
Quantity in commerce
47 units
Distribution
US nationwide in the states of ME, LA, and TN
Product codes and lots
Lot: 17971870

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1370-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1370-2016Class IITerminatedVoluntary: Firm initiated