Z-1533-2016Class I

Dialog+ Hemodialysis machines: Dialog + 120 V DF+Adimea+WAN-BSL Product Code # 710200S Serial # >200805 and...

B Braun Medical Inc / initiated 2016-04-01 / Terminated / root cause: manufacturing process

B Braun Medical Inc began a recall of Dialog+ Hemodialysis machines: Dialog + 120 V DF+Adimea+WAN-BSL Product Code # 710200S Serial # >200805 and <204001 The Dialog+ can be used for implementing... on . FDA classified it as Class I and gives the reason as "Potential leakage of conductivity sensors of the Dialog Dialysis machines". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1533-2016
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B Braun Medical Inc
Product
Dialog+ Hemodialysis machines: Dialog + 120 V DF+Adimea+WAN-BSL Product Code # 710200S Serial # >200805 and <204001 The Dialog+ can be used for implementing and monitoring hemodialysis treatments for patients with acute or chronic kidney failure. The system can be used in hospital, health center and outpatient dialysis center settings when prescribed by a Physician. The following types of renal replacement therapy can be carried out with the system: Hemodialysis (HD) with or without phases of pure ultrafiltration, High flux hemodialysis and low flux hemodialysis.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
B. Braun mailed an Urgent Medical Device Correction notice to affected customers informing them of the Dialog+Hemodialysis System due to the potential for Bicarbonate and End conductivity cells (3456102A and 3456103A respectively) to exhibit cracks. Notice provided customers with actions for Qualified Service Technicians, Clinical Personnel Administering Dialysis treatment and adverse event reporting. All Dialog+Hemodialysis machines will be inspected at the customer's site by the customer Biomed, who has been trained and qualified by B. Braun Avitum. The inspection of each machine will be performed in accordance to a Field Service Information (FSI). Customers with questions were instructed to contact the Technical Support Team at 800-621-0445 or the Customers Service Team at 800-848-2066. For questions regarding this recall call 610-997-4717.
Quantity in commerce
318 units
Distribution
Nationwide Distribution to ND, OH, CA, VA, FL, LA, SC, IL, UT, AZ, KY, IA, WI, OK, TN, SD, AL, NJ, TX, NM, NY, KS, WA, MO, CO, PA, CT, DE, GA, MD, VA, Puerto Rico, and St. Thomas USVI
Product codes and lots
Product Code 710200S Serial Number Range; 200805-204001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1533-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1531-2016Class ITerminatedVoluntary: Firm initiated
Z-1532-2016Class ITerminatedVoluntary: Firm initiated
Z-1533-2016Class ITerminatedVoluntary: Firm initiated
Z-1534-2016Class ITerminatedVoluntary: Firm initiated
Z-1535-2016Class ITerminatedVoluntary: Firm initiated
Z-1536-2016Class ITerminatedVoluntary: Firm initiated