Radionuclide Radiation Therapy System Product The product is a teletherapy device intended for stereotactic...
Elekta, Inc. began a recall of Radionuclide Radiation Therapy System Product The product is a teletherapy device intended for stereotactic irradiation of head structures ranging from very... on . FDA classified it as Class II and gives the reason as "The latches of the frame adapter can be locked even if the locating pins of the frame adapter is not inserted into the corresponding holes in the coordinate frame". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1561-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Radionuclide Radiation Therapy System Product The product is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.
- Product code
- IWB System, Radiation Therapy, Radionuclide
- Reason for recall
- The latches of the frame adapter can be locked even if the locating pins of the frame adapter is not inserted into the corresponding holes in the coordinate frame.
- FDA root cause
- Use error
- Action
- Elekta sent an Important User Notice (100-03-202-27) letter to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the Acknowledgement Form attached. For questions contact your local Elekta office.
- Quantity in commerce
- 217
- Distribution
- Worldwide Distribution - US Nationwide in the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WV, WI, and countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom
- Product codes and lots
- Perfexion - 14753-099/6052, 12579-007/6250, 10394-LGK01/6144, 11920-C5210/6088, 17397-001/6158, 105528-001/6122, 14857-001/6028, 12236-006/6097, 12473-003/6005, 14857-001/6028, 16867-001/6166, 12668-003/6145 10208-003/6215, 10698-001/6237, 10979-002/6213, 10316-500/6210, 21061-001/6195, 20364-001/6103, 11720-002/6143, 12620-002/6120, 10611-001/6081, 12846-004/6064, 11524-002/8001, 11420-001/6212, 11420-001/6228, 14983-001/6076, 11083-HOM001/6070, 10247-002/6136,10684-002/6090,11868-004/6002, 10795-K08002/6063, 20354-ENS/6160, 11792-005/6077, 11291-001/6018, 10326-003/6105, 18630-001/6031, 17322-001/6121, 30000656-002/6227, 12629-003/6026, 10924-001/6089, 11154-003/6050, 20438-003/6013 17522-001/6214, 17525-002/6038, 11414-LGK002/6205, 11940-LGK002/6181, 10514-LGK002/6183, 11639-LGK001/6042, 11390-LGK001/6109, 10416-LGK001/6125, 11670-LGK001/6133, 11990-LGK001/6118, 10648-LGK001/6098, 15104-LGK001/6080, 10654-LGK001/6096, 10959-LGK001/6132, 10997-LGK001/6087, 15748-LGK001/6106, 11149-LGK002/6206, 11986-LGK002/6218, 10929-LGK002/6175, 10879-LGK002/6234, 11434-LGK002/6255, 11991-LGK002/6198, 13480-LGK002/6148, 10425-LGK002/6254, 10583-LGK002/6161, 11563-LGK002/6162, 10103-LGK02/6149, 10541-LGK002/6142, 10848-LGK001/6108, 11433-LGK001/6116, 14900-LGK001/6058, 15045-LGK001/6091 11784-LGK001/6055, 30000732-001/6223, 40017-002/6170, 12085-006/6163, 12109-002/6169, 12959-002/6033, 10209-001/6046, 10960-005/6193, 0002718-001/6113, 30001751-WOJ701/6068, 12451-001/6019, 14878-003/6067, 105618-001/6110, 19634-001/6155, 16647-001/6187, 14891-001/6104, 30001527-001/6231, 12028-LGK002/6217, 11586-001/6157, 10728-001/6062, 11999-001/6126, 15416-001/6086, 10459-001/6123, 10158-001/6177, 10234-002/6111, 40023-001/6178, 10444-001/6085, 17033-003/6043, 11530-001/6101, 11553-001/6024, 10779-07/6069, 13465-CHUV03/6094, 10424-001/6182, 13370-002/6075, 12214-003/6007, 30000655-001/6199, 10152-001/6159, 11550-001/6065, 10118-001/6230, 12766-001/6112, 13160-001/6039, 11042-001/6003, 12270-001/6017, 18062-001/6146, 11710-004/6141, 11820-006/8006, 11570-004/6220, 11127-009/6219, 12164-004/6221, 10434-004/6244, 10542-012/6238, 13826-002/6239, 10567-041/6191, 98300-003/6140, 10567-037/6119, 14569-001/6135, 10643-005/6117, 11798-009/6060, 17357-001/6130, 11813-003/6034, 10356-007/6079, 10567-045/6232, 10567-032/6074, 10567-030/6036, 10567-025/6029, 11293-004/6004, 12269-002/6082, 11772-004/6048, 10164-004/6009, 10349-003/6066, 10567-026/6011, 10560-003/6127, 11073-008/6020, 10567-036/6102, 14236-001/6114, 10567-035/6092, 10379-002/6061, 11030-006/6030, 13050-001/6037, 10909-009/6010, 11714-003/6072, 12985-003/6025, 14030-002/6015, 10567-031/6054, 10064-008/6021, 13558-001/6099, 10567-038/6107, 11939-003/6008, 10591-002/6083, 10567-024/6014, 11848-003/6095, 11624-003/6078, 10383-007/6100, 11395-010/6032, 11198-005/6124, 10982-003/6053, 11256-001/6012, 10905-004/6016, 10387-011/6093, 11138-004/6115, 10524-003/6073, 10908-003/6045, 10762-005/6057, 10193-007/6056, 10673-016/6006, 11007-005/6035, 11791-006/6022, 11377-003/6186, 12752-004/6242, 11039-006/6188, 10597-002/6167, 17539-001/6185, 14153-001/6197, 30000118-001/6240, 10567-039/6180, 15053-001/6179, 11731-005/6200, 11090-005/6165, 11077-007/6207, 30000171-002/6172, 10760-006/6173, 11733-001/6192, 10817-009/6204, 13514-001/6211, 10877-GK01/6151, 10436-009/6150, 11396-003/6176, 10917-004/6128, 11373-002/6184, 10567-042/6129, 10567-043/6203, 12614-003/6139, 10297-006/6171, 11347-003/6134 LEKSELL GAMMA KNIFE ICON - 11868-008/6131, 11290-ENS1/6174, 12027-EZT-G1/6041, 10444-LGK004/8003, 11352-001/6044, 10622-005/6194, 10886-013/6027, 11201-004/6047, 11834-003/6222
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1561-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY