Z-1506-2016Class II

IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH...

Philips North America LLC / initiated 2016-03-04 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides software applications used to process... on . FDA classified it as Class II and gives the reason as "The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1506-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.
Root cause tag
software
FDA root cause
Software design
Action
On March 4, 2016, the firm sent an URGENT - Field Safety Notice Medical Device Correction letter to their customers. The letter identified the affected device, the reason for the recall, and under what circumstances the problem occurs. Customers are to contact Philips as described if they are unsure whether their gamma cameras supports the modes. The firm will implement the AutoSPECT Pro software upgrade and contact customers to schedule the appointment. If further assistance or support is needed, customers are to contact their local Philips representative.
Quantity in commerce
33 Units
Distribution
Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.
Product codes and lots
M/N 881001; S/N: 85254, 82202, 85330, 82117, 7000022, 105009, 970001, 850001, 870030, 82141, 95068, 87182, 87197, 95078, 82140, 11017, 98139, 95035, 87200, 980047, 85314, 85303, 980046, 85298, 822022, 7000019, 950076, 950115, 95043, 95047, 950021, 960123 & 105006.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1506-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1504-2016Class IITerminatedVoluntary: Firm initiated
Z-1505-2016Class IITerminatedVoluntary: Firm initiated
Z-1506-2016Class IITerminatedVoluntary: Firm initiated