Z-1833-2016Class II

The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-wide...

Abbott Medical / initiated 2016-03-25 / Terminated / root cause: manufacturing process

Hospira Inc. began a recall of The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare... on . FDA classified it as Class II and gives the reason as "Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1833-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With its primary line, secondary line, and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories. Each system includes a pumping module (hereafter called the infuser) and an assortment of disposable IV sets (hereafter called a set), and optional accessories. The Plum A+ host device contains a Connectivity Engine peripheral module that provides wired Ethernet and wireless 802.11 a/b/g local area networking capabilities. This allows the Hospira Mednet networked application software to download drug libraries to the infuser and enable the autoprogramming feature.
Product code
FRN Pump, Infusion
Reason for recall
Firm's investigation determined that the Proximal Occlusion Alarms observed in the field were attributed to the mechanism calibration process being performed with an out of specification metal cassette, resulting in an offset in proximal pressure.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Hospira Inc, sent an "URGENT MEDICAL DEVICE CORRECTION" letter on March 31, 2016, to the direct consignees of record via traceable mailing which included a data capture log containing a list of devices located with each customer. Delivery will be confirmed via traceable mailing. Consignees who further distributed the product were asked to notify their customers. Please take the following actions related to Urgent Medical Device Correction letter notification: 1. Inform potential users in your organization of this notification. 2. Complete the attached reply form and return it to the fax number or e-mail address on the form, even if you do not have the affected product. 3. If you have distributed the product further, notify your accounts that received the product identified above of this notification and ask them to contact Stericycle at 1-877-479-8088 (M-F, 8am-5pm ET) to receive a reply form. Hospira will contact you to make arrangements for your impacted Plum 360 I Plum A+ devices to be re-calibrated to correct this issue. :
Quantity in commerce
2567 Plum A+ infusion pumps.
Distribution
U.S. distribution : Natrionwide. Foreign :Brazil, Canada, Chile, Colombia, Ecuador, Guatemala, Mexico, Oman, Peru, Taiwan, Turkey, United Kingdom and Uruguay.
Product codes and lots
PLum A+ devices with the following list numbers: 12391-04-25, 12391-22-05, 12391-23-05, 12391-23-15, 12391-36-07, 12391-54-05, 12391-88-01, 12391-88-02, 12391-88-03, 12391-88-04, 12391-88-05, 20792-04-25, 20792-13-09, 20792-57-15, 20792-65-01, 20792-65-02 and 20792-88-69.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1833-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1832-2016Class IITerminatedVoluntary: Firm initiated
Z-1833-2016Class IITerminatedVoluntary: Firm initiated