Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1593-2016Class II

CARESTREAM Image Suite V3: MINI-PACS/F IMG ST/CLASSIC: REF/Catalog # 1036490; MINI-PACS/F IMG ST/POC&VITA...

Eastman Kodak Company / initiated 2016-04-15 / Terminated / root cause: software

Carestream Health Inc began a recall of CARESTREAM Image Suite V3: MINI-PACS/F IMG ST/CLASSIC: REF/Catalog # 1036490; MINI-PACS/F IMG ST/POC&VITA: REF/Catalog # 1036508; MINI-PACS/F IMG... on . FDA classified it as Class II and gives the reason as "Carestream Health received a complaint related to CARESTREAM Image Suite 4 from a foreign hospital stating that the annotation on the overlay is displayed as "L (Left)", when it should be "R(Right)"". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1593-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Eastman Kodak Company
Product
CARESTREAM Image Suite V3: MINI-PACS/F IMG ST/CLASSIC: REF/Catalog # 1036490; MINI-PACS/F IMG ST/POC&VITA: REF/Catalog # 1036508; MINI-PACS/F IMG ST/CLASSIC/INDIA: REF/Catalog # 1036417; MINI-PACS/F IMG ST/ POC&VITA /INDIA: REF/Catalog # 1036425 -- Made in USA by: Carestream Health, Inc. 150 Verona Street, Rochester, NY 14608 --- CLASSIFICATION NAME: System, Image Processing, Radiological The Carestream Image Suite System is an image management system whose intended use is to receive, process, review, display, print and archive images and data from CR and DR modalities. This excludes mammography applications in the United States.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Carestream Health received a complaint related to CARESTREAM Image Suite 4 from a foreign hospital stating that the annotation on the overlay is displayed as "L (Left)", when it should be "R(Right)".
Root cause tag
software
FDA root cause
Software design
Action
Carestream Health sent an URGENT MEDICAL DEVICE CORRECTION letter dated April 6, 2016, and Consignee Notification Acknowledgement form to the US consignees on April 15, 2016, via Certified Mail, Return Receipt Requested. Foreign consignees are being notified based on the individual country regulations. Customers who have this Image Suite configuration but do not use it for CT or MR Imaging do not need to take any action. Customers who use the product for CT or MR Images are asked to make users aware of the problem and take additional steps to check correct orientation of any markers, until the software is updated. Customers with questions or concerns were instructed to contact the Carestream Customer Care Center in the US at 1-800-328-2910. Customers outside the US were instructed to contact their local Service Support number. For questions regarding this recall call 1-800-328-2917.
Quantity in commerce
US: 50 units; Foreign: 2499 units
Distribution
NY and OH for further distribution nationwide; Foreign distribution to the following countries: Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzgovena, Brazil, Bulgaria, Cambodia, Canada, Chad, Chile, China, Colombia, Croatia, Czech Rep., Djibouti, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon, Liechtenstein, Lithuania, Macao, Malaysia, Mexico, Montenegro, Myanmar (Burma), Nepal, Netherlands, New Zealand, Niger, Norway, Oman, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Scotland, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Taiwan, thailand, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam.
Product codes and lots
Equipment ID Numbers: US units: 8501-2596, 85018860, 8501-0954, 5241-2596, 5241-2546, 5241-2713, 5296-3447, 52414419, 52414455, 5241-2640, 5241-2716, 54838552, 5483-8502, 8501-4665, 52408158, 85014687, 54839128, 5483-9212, 52412539, 52414661, 52962897, 4745-3986, 5241-2337, 85012262, 8501-5898, 8501-2602, 54838630, 52412670, 5296-3450, 5296-3449, 54838554, 52408157, 54838618, 5241-4547, 54842652, 52420675, 8501-2861,  52414685, 52414751, 5484-3086, 54838598, 52412367, 52414417, 54839211, 54838616, 4745-5991, 52412243, 54842667, 52417149, 52412340; -- Equipment ID Numbers for Foreign Units: Please contact CDRH Recall Group for list of affected Equipment ID Numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1593-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1593-2016Class IITerminatedVoluntary: Firm initiated
Z-1594-2016Class IITerminatedVoluntary: Firm initiated