Novocastr Liquid Mouse Monoclonal Antibody Renal Cell Carcinoma Marker (RCC) is recommended for use as part...
Leica Microsystems, Inc. began a recall of Novocastr Liquid Mouse Monoclonal Antibody Renal Cell Carcinoma Marker (RCC) is recommended for use as part of a panel of antibodies in the definition of... on . FDA classified it as Class II and gives the reason as "Testing of retained units indicated that Lot numbers of CDX2 and RCC indicated are not stable up to the expiry date on the product labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1876-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Leica Microsystems, Inc.
- Product
- Novocastr Liquid Mouse Monoclonal Antibody Renal Cell Carcinoma Marker (RCC) is recommended for use as part of a panel of antibodies in the definition of tumors of proximal nephrogenic differentiation.
- Product code
- NJT Immunohistochemistry Reagents And Kits
- Reason for recall
- Testing of retained units indicated that Lot numbers of CDX2 and RCC indicated are not stable up to the expiry date on the product labeling.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- An Urgent: Medical Device Recall Notification dated 4/5/2016 was sent to consignees with the affected lots of CDX2 and RCC marker. The customers were notified of the problems as well as the following actions to be taken by the consignees: Do not use or continue to use the reagent lots listed above, as they have the potential to result in inadequate tissue staining and lead to weak or false negative pathology results. Appropriately destroy any unused or partially used affected lots of the reagent Complete the attached Recall Notification Acknowledgement Form that this action has been undertaken. The firm also request that the notification is passed to all those within their organization who need to be aware of this issue. The firm requests consignees to return the completed acknowledgement form to [email protected] to receive credit for affected lots that have been destroyed by the consignees laboratory.
- Quantity in commerce
- 270 units
- Distribution
- Distributed in the following states AR, CA, CT, FL, ID, IL, IN, MT, OH, PA, TN and TX, and in Canada.
- Product codes and lots
- 1ml size Novocastr Liquid Mouse Monoclonal Antibody Renal Cell Carcinoma Marker (RCC) with lot numbers 6036120, 6032345 and 6023543.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1876-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1877-2016 | Class II | Terminated | Voluntary: Firm initiated |