Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in...
Leica Microsystems, Inc. began a recall of Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues on . FDA classified it as Class II and gives the reason as "Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1877-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Leica Microsystems, Inc.
- Product
- Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues.
- Product code
- NJT Immunohistochemistry Reagents And Kits
- Reason for recall
- Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- An Urgent: Medical Device Recall Notification dated 4/5/2016 was sent to consignees with the affected lots of CDX2 and RCC marker. The customers were notified of the problems as well as the following actions to be taken by the consignees: Do not use or continue to use the reagent lots listed above, as they have the potential to result in inadequate tissue staining and lead to weak or false negative pathology results. Appropriately destroy any unused or partially used affected lots of the reagent Complete the attached Recall Notification Acknowledgement Form that this action has been undertaken. The firm also request that the notification is passed to all those within their organization who need to be aware of this issue. The firm requests consignees to return the completed acknowledgement form to [email protected] to receive credit for affected lots that have been destroyed by the consignees laboratory.
- Quantity in commerce
- 90 Units
- Distribution
- Distributed in the following states AR, CA, CT, FL, ID, IL, IN, MT, OH, PA, TN and TX, and in Canada.
- Product codes and lots
- 1ml size Novocastr Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) with lot numbers 6032734, 6033766, 6037153 and 6039088.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1876-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1877-2016 | Class II | Terminated | Voluntary: Firm initiated |