Z-1877-2016Class II

Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in...

Leica Microsystems, Inc. / initiated 2016-04-05 / Terminated / root cause: manufacturing process

Leica Microsystems, Inc. began a recall of Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues on . FDA classified it as Class II and gives the reason as "Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1877-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Microsystems, Inc.
Product
Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues.
Product code
NJT Immunohistochemistry Reagents And Kits
Reason for recall
Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An Urgent: Medical Device Recall Notification dated 4/5/2016 was sent to consignees with the affected lots of CDX2 and RCC marker. The customers were notified of the problems as well as the following actions to be taken by the consignees: Do not use or continue to use the reagent lots listed above, as they have the potential to result in inadequate tissue staining and lead to weak or false negative pathology results. Appropriately destroy any unused or partially used affected lots of the reagent Complete the attached Recall Notification Acknowledgement Form that this action has been undertaken. The firm also request that the notification is passed to all those within their organization who need to be aware of this issue. The firm requests consignees to return the completed acknowledgement form to [email protected] to receive credit for affected lots that have been destroyed by the consignees laboratory.
Quantity in commerce
90 Units
Distribution
Distributed in the following states AR, CA, CT, FL, ID, IL, IN, MT, OH, PA, TN and TX, and in Canada.
Product codes and lots
1ml size Novocastr Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) with lot numbers 6032734, 6033766, 6037153 and 6039088.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1877-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1876-2016Class IITerminatedVoluntary: Firm initiated
Z-1877-2016Class IITerminatedVoluntary: Firm initiated