Vortex MEDICAL AngioVac Circuit, REF/UPN VTX3100, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit...
Angiodynamics Inc. (Navilyst Medical Inc.) began a recall of Vortex MEDICAL AngioVac Circuit, REF/UPN VTX3100, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit Pack are designed to be used... on . FDA classified it as Class II and gives the reason as "Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1701-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Vortex MEDICAL AngioVac Circuit, REF/UPN VTX3100, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit Pack are designed to be used with the AngioVac Cannula.
- Product code
- DWE Tubing, Pump, Cardiopulmonary Bypass
- Reason for recall
- Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- On 3/21/2016, AngioDynamics sent recall notification packages (dated 3/21/2016) to the consignees via Federal Express. Consignees are instructed to follow the instructions provided on the Maquet Cardiopulmonary recall letter (dated 2/23/2016) and return the Reply Verification Tracking Form, provided in the recall notification.
- Quantity in commerce
- Domestic: 339 kits
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lot Numbers 37FE2901, 37GD2908, 37KD2405
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1701-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1702-2016 | Class II | Terminated | Voluntary: Firm initiated |