AngioVac Circuit Pack (Tandem Packs), REF/Catalog No. 25-187, UPN H965251870, STERILE, Rx Only, Single Use...
Angiodynamics Inc. (Navilyst Medical Inc.) began a recall of AngioVac Circuit Pack (Tandem Packs), REF/Catalog No. 25-187, UPN H965251870, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit... on . FDA classified it as Class II and gives the reason as "Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1702-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- AngioVac Circuit Pack (Tandem Packs), REF/Catalog No. 25-187, UPN H965251870, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit Pack are designed to be used with the AngioVac Cannula.
- Product code
- DWE Tubing, Pump, Cardiopulmonary Bypass
- Reason for recall
- Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- On 3/21/2016, AngioDynamics sent recall notification packages (dated 3/21/2016) to the consignees via Federal Express. Consignees are instructed to follow the instructions provided on the Maquet Cardiopulmonary recall letter (dated 2/23/2016) and return the Reply Verification Tracking Form, provided in the recall notification.
- Quantity in commerce
- Domestic: 377 kits
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lot Numbers: 4876386, 4889699, 4889700, 4898652, 4902111, 4902117, 4906576, 4941511, 4941511D, 4943459, 4955686, 4956929, 4969475, 4975493, 4991710, 4999714.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1701-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1702-2016 | Class II | Terminated | Voluntary: Firm initiated |