Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating...
C.R. Bard, Inc. began a recall of Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating system indicated for monitoring and controlling... on . FDA classified it as Class II and gives the reason as "The product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM). This...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1617-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- C.R. Bard Inc
- Product
- Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating system indicated for monitoring and controlling patient temperature
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- The product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM). This premature coin cell battery drainage could render the device unresponsive upon System startup.
- FDA root cause
- Unknown/Undetermined by firm
- Action
- BARD sent an Urgent Medical Product Recall to all affected consignees on November 9, 2015. The letter indicates that a technician will contract the firm for the evaluation and potential replacement of the affected electronic components. The technician will fill out the Field Action Effectiveness Check Form after evaluation. Customers with questions were instructed to call 1-770-784-6471 or email [email protected]. For questions regarding this recall call 770-784-6220.
- Quantity in commerce
- 2613
- Distribution
- Worldwide Distribution to AL, AK, AZ, AR, CA, CO, CT, DEL, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and Internationally to Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Finland, France, Germany, Gibraltar, Greece, Guam, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom
- Product codes and lots
- Product Codes - 5000-00-00, 5000-01-01, 5000-01-02, 5000-01-03, 5000-01-04, 5000-01-05
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1617-2016 | Class II | Terminated | Voluntary: Firm initiated |