Z-1617-2016Class II

Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating...

C.R. Bard Inc / initiated 2015-11-09 / Terminated

C.R. Bard, Inc. began a recall of Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating system indicated for monitoring and controlling... on . FDA classified it as Class II and gives the reason as "The product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM). This...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1617-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.R. Bard Inc
Product
Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating system indicated for monitoring and controlling patient temperature
Product code
DWJ System, Thermal Regulating
Reason for recall
The product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM). This premature coin cell battery drainage could render the device unresponsive upon System startup.
FDA root cause
Unknown/Undetermined by firm
Action
BARD sent an Urgent Medical Product Recall to all affected consignees on November 9, 2015. The letter indicates that a technician will contract the firm for the evaluation and potential replacement of the affected electronic components. The technician will fill out the Field Action Effectiveness Check Form after evaluation. Customers with questions were instructed to call 1-770-784-6471 or email [email protected]. For questions regarding this recall call 770-784-6220.
Quantity in commerce
2613
Distribution
Worldwide Distribution to AL, AK, AZ, AR, CA, CO, CT, DEL, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and Internationally to Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Finland, France, Germany, Gibraltar, Greece, Guam, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Oman, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom
Product codes and lots
Product Codes - 5000-00-00, 5000-01-01, 5000-01-02, 5000-01-03, 5000-01-04, 5000-01-05