Z-1558-2016Class II

SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of...

Siemens Medical Solutions USA, Inc. / initiated 2016-04-01 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray... on . FDA classified it as Class II and gives the reason as "The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1558-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or liquid is not depicted as expected). There is also a risk of a potential misdiagnosis in using this protocol for surgery planning. Investigations revealed an incorrect parameterization of the reconstruction algorithm applied for the pediatric head kernel as cause of the problem.
Root cause tag
software
FDA root cause
Software design
Action
Siemens distributed a Safety Advisory Notice dated April 1, 2016, to affected customers The letter identified the affected product, problem, how to avoid potential risks, how the issue will be resolved. and actions to be taken. Customers were asked to observe the safety notice and comply with the corresponding measures until further notice.
Quantity in commerce
37 CTsystems
Distribution
US Nationwide Distribution : NC, NH, IL, MA, WA, PA, MD, MN, TX, NY, SC, OH, CA, KY, MT, IA
Product codes and lots
SOMATOM Force Material # 10742326, serial numbers 75550, 75526, 75512, 75567, 75536, 75514, 75563, 75510, 75454, 75540, 75571, 75437, 75500, 75527, 75475, 75535, 75481, 75439, 75450, 75482, 75460, 75493, 75476, 75458, 75528, 75467, 75487, 75478, 75513, 75532, 75555, 75524, 75515, 75559, 75501, 75570, 75576, 75450, 75482, 75528, 75576

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1558-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1558-2016Class IITerminatedVoluntary: Firm initiated