SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of...
Siemens Medical Solutions USA, Inc began a recall of SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray... on . FDA classified it as Class II and gives the reason as "The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1558-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- SOMATOM Force, Computed Tomography x-ray system intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or liquid is not depicted as expected). There is also a risk of a potential misdiagnosis in using this protocol for surgery planning. Investigations revealed an incorrect parameterization of the reconstruction algorithm applied for the pediatric head kernel as cause of the problem.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens distributed a Safety Advisory Notice dated April 1, 2016, to affected customers The letter identified the affected product, problem, how to avoid potential risks, how the issue will be resolved. and actions to be taken. Customers were asked to observe the safety notice and comply with the corresponding measures until further notice.
- Quantity in commerce
- 37 CTsystems
- Distribution
- US Nationwide Distribution : NC, NH, IL, MA, WA, PA, MD, MN, TX, NY, SC, OH, CA, KY, MT, IA
- Product codes and lots
- SOMATOM Force Material # 10742326, serial numbers 75550, 75526, 75512, 75567, 75536, 75514, 75563, 75510, 75454, 75540, 75571, 75437, 75500, 75527, 75475, 75535, 75481, 75439, 75450, 75482, 75460, 75493, 75476, 75458, 75528, 75467, 75487, 75478, 75513, 75532, 75555, 75524, 75515, 75559, 75501, 75570, 75576, 75450, 75482, 75528, 75576
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1558-2016 | Class II | Terminated | Voluntary: Firm initiated |