Z-1634-2016Class II

CinchLock SS Knotless Anchor with Inserter; Model number CAT02462; Orthopedic: The Pivot CinchLock SS...

Stryker Corp. / initiated 2016-04-12 / Terminated / root cause: design

Stryker Corporation began a recall of CinchLock SS Knotless Anchor with Inserter; Model number CAT02462; Orthopedic: The Pivot CinchLock SS Knotless Anchor (CAT02462) and Pivot CinchLock Flex... on . FDA classified it as Class II and gives the reason as "Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of s of an internal wire breaking upon deployment of the anchor". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1634-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
CinchLock SS Knotless Anchor with Inserter; Model number CAT02462; Orthopedic: The Pivot CinchLock SS Knotless Anchor (CAT02462) and Pivot CinchLock Flex Knotless Anchor (CAT02463) are non-degradable implant devices intended for the fixation of soft tissue to bone. Used for repair of hip, shoulder, foot and ankle, elbow, wrist and hand and knee.
Product code
MBI Fastener, Fixation, Nondegradable, Soft Tissue
Reason for recall
Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of s of an internal wire breaking upon deployment of the anchor.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Medical Device Recalls letters were sent on April 12, 2016 to US customers by certified mail. Stryker International distribution sites will be notified by e-mail. Letters describe the reason for recall with the affected part and lot numbers. Any recalled product should be segregated and returned to Stryker.
Quantity in commerce
3100 total device, both model numbers.
Distribution
Distributed US (nationwide) and in Australia, Switzerland, Netherlands, Spain and Sweden.
Product codes and lots
Model number: CAT02462 Serial numbers: 15120801, 15121401, 16010402, 16011901, 16012501.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1634-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1634-2016Class IITerminatedVoluntary: Firm initiated