CinchLock Flex Knotless Anchor with Inserter; Model number CAT02643; Orthopedic: The Pivot CinchLock SS...
Stryker Corporation began a recall of CinchLock Flex Knotless Anchor with Inserter; Model number CAT02643; Orthopedic: The Pivot CinchLock SS Knotless Anchor (CAT02462) and Pivot CinchLock Flex... on . FDA classified it as Class II and gives the reason as "Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of an internal wire breaking upon deployment of the anchor". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1635-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- CinchLock Flex Knotless Anchor with Inserter; Model number CAT02643; Orthopedic: The Pivot CinchLock SS Knotless Anchor (CAT02462) and Pivot CinchLock Flex Knotless Anchor (CAT02463) are non-degradable implant devices intended for the fixation of soft tissue to bone. Used for repair of hip, shoulder, foot and ankle, elbow, wrist and hand and knee.
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Stryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of an internal wire breaking upon deployment of the anchor.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Urgent Medical Device Recalls letters were sent on April 12, 2016 to US customers by certified mail. Stryker International distribution sites will be notified by e-mail. Letters describe the reason for recall with the affected part and lot numbers. Any recalled product should be segregated and returned to Stryker.
- Quantity in commerce
- 3100 total devices, both models
- Distribution
- Distributed US (nationwide) and in Australia, Switzerland, Netherlands, Spain and Sweden.
- Product codes and lots
- Model number: CAT02643; Serial numbers: 15122202, 16011301,16020802
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1634-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1635-2016 | Class II | Terminated | Voluntary: Firm initiated |