Z-1572-2016Class II

ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844 ACUSON X700 Ultrasound...

Siemens Medical Solutions USA, Inc. / initiated 2015-09-05 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844 ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System... on . FDA classified it as Class II and gives the reason as "Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1572-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844 ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System (Russia) 10658845 - ACUSON X700 Ultrasound System (Korea) Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesphageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which requires a reboot to recover system operation.
Root cause tag
software
FDA root cause
Software design
Action
September 5, 2015 we initiated a correction via update program US020/15/S where we visited all four units and updated their software to version 1.0.05
Quantity in commerce
4
Distribution
Distributed in Germany, U.A.E., Hungary and Brazil.
Product codes and lots
Material (System) IVK Group Serial number: 10658844 ACUSON X700 356559 10658844 ACUSON X700 356733 10658844 ACUSON X700 356734 10658844 ACUSON X700 358714

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1572-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1572-2016Class IITerminatedVoluntary: Firm initiated