Z-1601-2016Class II

Syngo Dynamics; Kinetdx Picture Archiving and Communications System

Siemens Medical Solutions USA, Inc. / initiated 2016-04-14 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo Dynamics; Kinetdx Picture Archiving and Communications System on . FDA classified it as Class II and gives the reason as "Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1601-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo Dynamics; Kinetdx Picture Archiving and Communications System
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.
Root cause tag
software
FDA root cause
Software design
Action
Siemens mailed out a "Customer Advisory Notice" #SY009/16/P to customers informing them of the issue and what to do if it was to occur. Siemens informed customers that they are working on the "carry forward" feature enhancement and is planning to release a software update VA10B_HF06 and they should follow the instructions in the advisory notice until the update is performed.
Quantity in commerce
84 units
Distribution
Nationwide: AZ, CA, IL, IN, MA, MD, MI, MN, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI
Product codes and lots
Model Numbers: 10091637, 10091673, 10091805, 10091807  Serial Numbers: 85021, 85040, 85042, 85045, 85047, 85057, 85079, 85080, 85081, 85086, 85090, 85115, 85143, 85145, 85146, 85147, 85149, 85157, 85167, 85168, 85170, 85172, 85177, 85178, 85185, 85186, 85189, 85191, 85234, 85235, 85239, 85281, 85307, 85309, 85322, 85325, 85326, 85329, 85335, 85338, 85350, 85353, 85354, 85368, 85370, 85374, 85377, 85380, 85385, 85387, 85401, 85418, 85421, 85422, 85427, 85435, 85442, 85447, 85462, 85467, 85468, 85469, 85470, 85473, 85481, 85482, 85484, 85859, 85860, 85866, 85867, 85871, 85872, 85873, 85874, 85875, 85876, 85880, 85883, 85884, 85893, 85896, 85910, 85918

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1601-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1601-2016Class IITerminatedVoluntary: Firm initiated