Syngo Dynamics; Kinetdx Picture Archiving and Communications System
Siemens Medical Solutions USA, Inc began a recall of Syngo Dynamics; Kinetdx Picture Archiving and Communications System on . FDA classified it as Class II and gives the reason as "Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1601-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Syngo Dynamics; Kinetdx Picture Archiving and Communications System
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens mailed out a "Customer Advisory Notice" #SY009/16/P to customers informing them of the issue and what to do if it was to occur. Siemens informed customers that they are working on the "carry forward" feature enhancement and is planning to release a software update VA10B_HF06 and they should follow the instructions in the advisory notice until the update is performed.
- Quantity in commerce
- 84 units
- Distribution
- Nationwide: AZ, CA, IL, IN, MA, MD, MI, MN, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI
- Product codes and lots
- Model Numbers: 10091637, 10091673, 10091805, 10091807 Serial Numbers: 85021, 85040, 85042, 85045, 85047, 85057, 85079, 85080, 85081, 85086, 85090, 85115, 85143, 85145, 85146, 85147, 85149, 85157, 85167, 85168, 85170, 85172, 85177, 85178, 85185, 85186, 85189, 85191, 85234, 85235, 85239, 85281, 85307, 85309, 85322, 85325, 85326, 85329, 85335, 85338, 85350, 85353, 85354, 85368, 85370, 85374, 85377, 85380, 85385, 85387, 85401, 85418, 85421, 85422, 85427, 85435, 85442, 85447, 85462, 85467, 85468, 85469, 85470, 85473, 85481, 85482, 85484, 85859, 85860, 85866, 85867, 85871, 85872, 85873, 85874, 85875, 85876, 85880, 85883, 85884, 85893, 85896, 85910, 85918
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1601-2016 | Class II | Terminated | Voluntary: Firm initiated |