510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1709-2016Class II

3.2 mm Three-fluted drill bit, sterile, Catalog ID # 03.010.060S

Synthes (USA) Products LLC / initiated 2016-04-18 / Terminated / root cause: design

Synthes (USA) Products LLC began a recall of 3.2 mm Three-fluted drill bit, sterile, Catalog ID # 03.010.060S Product Usage: Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with... on . FDA classified it as Class II and gives the reason as "The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1709-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
3.2 mm Three-fluted drill bit, sterile, Catalog ID # 03.010.060S Product Usage: Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with the following systems: Expert Lateral Femoral Nail (part number 03.010.060S only), Expert Tibial Nail, Expert Humeral Nailing, Multiloc Humeral Nailing System, Suprapatellar Instrumentation for Expert Tibial Nail, and Adolescent Lateral Femoral Nails (ALFN).
Product code
HTW Bit, Drill
Reason for recall
The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.
Root cause tag
design
FDA root cause
Device Design
Action
DePuy Synthes sent an Urgent Notice Medical Device Letter dated April 19, 2016 to customers. The identified the affected product, problem and actions to be taken. The letter instructed customers to examine their inventory to determine if they currently have the affected product. Customers were asked to complete and return the Verification Section of the recall notification letter. For questions call 610-719-5450.
Quantity in commerce
7363 units
Distribution
US Nationwide distribution in the states of: AE, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,UT, VA, VT, WA, WI
Product codes and lots
Catalog ID 03.010.060S Lot # 5558809, 6204324, 6813586, 7648837, 5559839, 6214366, 6818438, 7648838, 5575915, 6235676, 6827807, 7657752, 5579717, 6240287, 6830470, 7657753, 5582298, 6247514, 6835634, 7660380, 5585788, 6249834, 6841962, 7661528 5594237, 6269336, 6851101, 7661529, 5594238, 6276953, 6857546, 7674414, 5616439, 6279756, 6860704, 7687319, 5633067, 6279757, 6870563, 7693164, 5656914, 6302746, 6873394, 7716604, 5672506, 6313568, 6873396, 7731690 5682043, 6313569, 6898040, 7739027, 5692087, 6325420, 6904377, 7753584, 5702206, 6331923, 6908328, 7766654, 5711966, 6334954, 6919278, 7774127, 5724637, 6339936, 6925232, 7774128, 5727691, 6348925, 6931055, 7794479 5732158, 6355191, 6939080, 7817283, 5741332, 6368986, 6949503, 7830606, 5748361, 6375334, 6961056, 7835732, 5760640, 6382329, 6967764, 7848333, 5772367, 6398746, 6975982, 7869029, 5780654, 6413506, 6984973, 7870629 5782800, 6420887, 6996252, 7894923, 5782801, 6433683, 7004816, 7899079, 5782802, 6433684, 7013143, 7903010, 5794181, 6438144, 7019693, 7908522, 5816780, 6449832, 7029024, 7918012, 5818769, 6458079, 7036198, 7919639 5818770, 6464443, 7052563, 7924989, 5829138, 6475737, 7064072, 7941743, 5847359, 6480417, 7070681, 7947278, 5860702, 6486376, 7084937, 7963626, 5861834, 6488940, 7275546, 7981633, 5861836, 6505811, 7304754, 7993351 5870708, 6509276, 7349355, 9808655, 6010554, 6528479, 7359254, 9820617, 6031951, 6540263, 7371663, 9825386, 6034093, 6549146, 7382623, 9826363, 6034128, 6557658, 7400014, 9831036, 6051790, 6563112, 7400015, 9841499 6052762, 6572199, 7400016, 9855040, 6055036, 6577880, 7416878, 9859105, 6066429, 6586607, 7422660, 9864677, 6069150, 6593553, 7441691, 9869742, 6069151, 6598206, 7444103, 9878216, 6074684, 6603184, 7460590, 9886333 6080184, 6610704, 7465210, 9904888, 6081960, 6621162, 7476937, 9911126, 6091446, 6636693, 7485616, 9915461, 6092777, 6655352, 7497778, 9924572, 6109266, 6674053, 7499940, 9932287, 6121870, 6686555, 7512119, 9939768 6124816, 6707678, 7520264, 9959608, 6130838, 6721722, 7533286, 9961574, 6133623, 6733633, 7544212, 9970453, 6136836, 6752258, 7554690, 9985425, 6141455, 6761019, 7563621, H023093, 6153365, 6766654, 7636777, H024812 6163583, 6777654, 7636778, H034525, 6186051, 6786129, 7636779, H039109, 6186052, 6796018, 7648835, 6187891, 6796019, 7648836

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1709-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1709-2016Class IITerminatedVoluntary: Firm initiated
Z-1710-2016Class IITerminatedVoluntary: Firm initiated