3.2 mm three-fluted Radiolucent Drill Bit/Needle Point/145 mm Sterile
Synthes (USA) Products LLC began a recall of 3.2 mm three-fluted Radiolucent Drill Bit/Needle Point/145 mm Sterile Product Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with... on . FDA classified it as Class II and gives the reason as "The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1710-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes (USA) Products LLC
- Product
- 3.2 mm three-fluted Radiolucent Drill Bit/Needle Point/145 mm Sterile Product Usage: The 3.2 mm Three-fluted Drill Bits are instruments that can be used with the following systems: Expert Lateral Femoral Nail (part number 03.010.060S only), Expert Tibial Nail, Expert Humeral Nailing, Multiloc Humeral Nailing System, Suprapatellar Instrumentation for Expert Tibial Nail, and Adolescent Lateral Femoral Nails (ALFN).
- Product code
- HTW Bit, Drill
- Reason for recall
- The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- DePuy Synthes sent an Urgent Notice Medical Device Letter dated April 19, 2016 to customers. The identified the affected product, problem and actions to be taken. The letter instructed customers to examine their inventory to determine if they currently have the affected product. Customers were asked to complete and return the Verification Section of the recall notification letter. For questions call 610-719-5450.
- Quantity in commerce
- 1274 units
- Distribution
- US Nationwide distribution in the states of: AE, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,UT, VA, VT, WA, WI
- Product codes and lots
- Catalog ID 03.010.100S Lot # 5525872, 6153368, 6530901, 5548707, 6197315, 6581051 5575918, 6199232, 6676406, 5603463, 6231850, 6721724 5649117, 6256560, 6734876, 5668880, 6305842, 6808463 5724639, 6319050, 6831468, 5742276, 6320150, 6997294 5807341, 6343980, 7039877, 6012770, 6401819, 7052565 6045597, 6448707, 7081479, 6057914, 6472100, 7378832 6108342, 6501856, 7410311, 6142443, 6503351, 7466207 7554670, 7620971, 7675571, 7711694, 7801251, 7918026 9815103, 9828932, 9905850, 9939769, H033348
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1709-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1710-2016 | Class II | Terminated | Voluntary: Firm initiated |