510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1598-2016Class II

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915...

Siemens Medical Solutions USA, Inc. / initiated 2016-04-13 / Terminated / root cause: supplier

Siemens Medical Solutions USA, Inc began a recall of MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of... on . FDA classified it as Class II and gives the reason as "The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1598-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
Siemens mailed to customers on April 13, 2016, a Customer Safety Advisory Notice, MR013/16/S, to inform them of the actions to take to prevent a possible hazard to persons and equipment. Customers were informed that they will be contacted by one of Siemens' service organization who will schedule an appointment to perform a correction, which is to exchange the top cover.
Quantity in commerce
8 systems
Distribution
US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.
Product codes and lots
Model # 10432914 and 10432915 Serial Numbers: 42472, 42489, 42477, 46169, 46178, 46175, 42478, 42485

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1598-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1598-2016Class IITerminatedVoluntary: Firm initiated