MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915...
Siemens Medical Solutions USA, Inc began a recall of MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of... on . FDA classified it as Class II and gives the reason as "The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1598-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- Siemens mailed to customers on April 13, 2016, a Customer Safety Advisory Notice, MR013/16/S, to inform them of the actions to take to prevent a possible hazard to persons and equipment. Customers were informed that they will be contacted by one of Siemens' service organization who will schedule an appointment to perform a correction, which is to exchange the top cover.
- Quantity in commerce
- 8 systems
- Distribution
- US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.
- Product codes and lots
- Model # 10432914 and 10432915 Serial Numbers: 42472, 42489, 42477, 46169, 46178, 46175, 42478, 42485
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1598-2016 | Class II | Terminated | Voluntary: Firm initiated |