510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1708-2016Class II

Reprocessed Daig Supreme Fixed Curve Diagnostic Electrophysiology Catheters; Model #: 401450, 401877...

Stryker Sustainability Solutions / initiated 2016-04-28 / Terminated / root cause: labeling and UDI

Stryker Sustainability Solutions began a recall of Reprocessed Daig Supreme Fixed Curve Diagnostic Electrophysiology Catheters; Model #: 401450, 401877. Indicated for temporary intracardiac sensing, recording... on . FDA classified it as Class II and gives the reason as "Mislabeled for size". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1708-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Sustainability Solutions
Product
Reprocessed Daig Supreme Fixed Curve Diagnostic Electrophysiology Catheters; Model #: 401450, 401877. Indicated for temporary intracardiac sensing, recording, stimulation, and electrophysiology mapping of cardiac structures.
Product code
NLH Catheter, Recording, Electrode, Reprocessed
Reason for recall
Mislabeled for size
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
The firm sent out customer notification letters on 04/28/16. Customers are instructed to discontinue the use of affected lots of recalled product. Other EP Catheters reprocessed by Stryker Sustainability Solutions should be considered as alternatives. A Recall Effectiveness Check Form including the quantity of affected devices shipped to facility is enclosed. Please complete the Recall Effectiveness Check Form and indicate if any devices from the affected lots remain in your inventory. This form needs to be completed even if no affected product is found. Please return the completed and signed Recall Effectiveness Check Form to your local Stryker Sustainability Sales Representative or email to [email protected] or mail to: Stryker Sustainability Solutions 1810 West Drake Drive Tempe, AZ, 85283 Attn: Jodie Rueckert If the form indicates that affected devices remain in inventory, a prepaid shipping label will be issued for the return of the product. Customers will receive a credit for all affected devices returned. Adverse reactions or quality problems experienced with the use of this product may be reported to: - Stryker Sustainability Solutions Complaint Hotline: +1(888) 888-3433 X5555 - http://www.stryker.com/productexperience/ - The FDAs MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. On 05/12/16 the firm sent out a clarification letter to emphasize that use of affected lots of product should be discontinued. In addition, SSS would like to note that, use of devices larger than expected may result in excess force applied to the blood vessel. The risk to the patient can vary from a prolongation in procedure time, vessel rupture, organ damage or other unforeseen consequences. Some of these injuries may require surgery or other procedures to remedy. Any questions contact the Stryker Sustainability Solutions Sales Representative or Regulatory Affairs.
Quantity in commerce
18 devices
Distribution
Distributed in the state of NY and the country of Canada.
Product codes and lots
Catalog Number: 401450 Lot Number: 2561683E Expiration Date: 5/13/2017  Catalog Number: 401877 Lot Number: 2555866E Expiration Date: 5/6/2017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1708-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1708-2016Class IITerminatedVoluntary: Firm initiated