Z-1996-2016Class II

EVARREST Fibrin Sealant Patch, Manufactured by: Omrix Biopharmaceuticals, Ltd, Israel. Indicated for use with...

Ethicon, Inc / initiated 2016-04-11 / Terminated / root cause: labeling and UDI

Ethicon, Inc. began a recall of EVARREST Fibrin Sealant Patch, Manufactured by: Omrix Biopharmaceuticals, Ltd, Israel. Indicated for use with manual compression as an adjunct to hemostasis... on . FDA classified it as Class II and gives the reason as "Ethicon is recalling EVARREST Fibrin Sealant Patch because the characters "0-bad" were printed in the expiration date field on the foil pouch labels of the Fibrin Patch units instead of the valid expiration date...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1996-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon, Inc
Product
EVARREST Fibrin Sealant Patch, Manufactured by: Omrix Biopharmaceuticals, Ltd, Israel. Indicated for use with manual compression as an adjunct to hemostasis for control of bleeding during adult liver surgery and soft tissue bleeding in adults when control of bleeding by standard surgical methods of hemostasis is ineffective or impractical.
Product code
GEL
Reason for recall
Ethicon is recalling EVARREST Fibrin Sealant Patch because the characters "0-bad" were printed in the expiration date field on the foil pouch labels of the Fibrin Patch units instead of the valid expiration date (2016-12-28).
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Ethicon sent an "Urgent: Product Recall (Removal)," dated April 14, 2016 via UPS next day mail to their affected customers. The letter identified the affected product and the reason for the recall. The letter also states that product from the affected lot is still within the expiration date, and health care practitioners, who have treated patients with the product should continue to follow patients in the usual manner. Users are to examine their inventory and quarantine the affected product. The affected product should be returned per the instructions outlined in the letter. Ethicon will provide replacement product upon receipt. Customers are to complete and return the Business Reply Form. If customers need assistance returning the product, they can contact Stericycle at 1-866-761-9472. If there are questions regarding the replacement product, customers can contact Ethicon Customer Support Center at 1-877-ETHICON.
Quantity in commerce
228 units
Distribution
Nationwide Distribution
Product codes and lots
Product Code #EVT5024, Lot #U01F012A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1996-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1996-2016Class IITerminatedVoluntary: Firm initiated