Z-2804-2016Class III

The VITROS Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS Chemistry Products...

Ortho-Clinical Diagnostics, Inc. / initiated 2016-04-19 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of The VITROS Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay... on . FDA classified it as Class III and gives the reason as "There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2804-2016
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
The VITROS Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.
Product code
JIT Calibrator, Secondary
Reason for recall
There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On 19 April 2016, Ortho Clinical Diagnostics distributed Correction Notification letters (Ref. CL2016-087) and ADD DRV 5877 notices to their US consignees via courier service and to their foreign affiliates via email. Customers were advised to install the Assay Data Diskette DRV 5877 or above on their VITROS system. Following a successful calibration and quality control assessment, it is acceptable to use VITROS HbA1c Reagent Kit, GEN 06. Customers who further distributed the product should also provide a copy of the enclosed letter to all customers who received the product. It is suggested that customers complete and return the Confirmation of Receipt form to indicate that you have been informed of this issue. Customers with questions can contact the Technical Solutions Center at 1-800-421-3311.
Quantity in commerce
520 Units
Distribution
Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.
Product codes and lots
Lot 3155 Product Codes: 6842906 & 6801876

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2804-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2804-2016Class IIITerminatedVoluntary: Firm initiated