Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in...
Cook Inc. began a recall of Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and... on . FDA classified it as Class II and gives the reason as "Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2613-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
- FDA root cause
- Unknown/Undetermined by firm
- Action
- Cook Inc. sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter dated April 15, 2016, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with any medical questions or concerns, were instructed to contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. For information regarding the recall, please contact Stericycle Expert Solutions at 866.201.9067.
- Quantity in commerce
- 60,302
- Distribution
- Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.
- Product codes and lots
- All Lots. Catalog prefixes/suffixes SCBR5.5-/-SHTL SCBR6.5-/-SHTL Catalog/product number; Global/order number SCBR5.5-35-125-P-NS-H1-SHTL G31129 SCBR5.5-35-125-P-NS-JB1-SHTL G31130 SCBR5.5-35-125-P-NS-JB2-SHTL G36361 SCBR5.5-35-125-P-NS-MPA-SHTL G23697 SCBR5.5-35-125-P-NS-SIM1-SHTL G23699 SCBR5.5-35-125-P-NS-SIM2-SHTL G36360 SCBR5.5-35-125-P-NS-VERT-SHTL G23698 SCBR5.5-35-125-P-NS-VTK-SHTL G36359 SCBR5.5-35-135-P-NS-ANG-SHTL G48154 SCBR5.5-35-150-P-NS-0-SHTL G23705 SCBR5.5-35-150-P-NS-ANG-SHTL G23704 SCBR5.5-35-75-P-NS-ANG-SHTL G48152 SCBR6.5-35-125-P-NS-H1-SHTL G31134 SCBR6.5-35-125-P-NS-JB1-SHTL G31135 SCBR6.5-35-125-P-NS-JB2-SHTL G36364 SCBR6.5-35-125-P-NS-SIM1-SHTL G23701 SCBR6.5-35-125-P-NS-SIM2-SHTL G36363 SCBR6.5-35-125-P-NS-VERT-SHTL G23700 SCBR6.5-35-125-P-NS-VTK-SHTL G36362
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2610-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2611-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2612-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2613-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2614-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2615-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2616-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2617-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2618-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2619-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2620-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2621-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2622-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2623-2016 | Class II | Terminated | Voluntary: Firm initiated |