Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2642-2016Class II

Ziehm Vision R, Mobile C-arm for Mobile Fluoroscopic Imaging

Ziehm Imaging GmbH / initiated 2016-04-15 / Terminated / root cause: design

Ziehm Imaging Inc began a recall of Ziehm Vision R, Mobile C-arm for Mobile Fluoroscopic Imaging on . FDA classified it as Class II and gives the reason as "Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2642-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ziehm Imaging GmbH
Product
Ziehm Vision R, Mobile C-arm for Mobile Fluoroscopic Imaging
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.
Root cause tag
design
FDA root cause
Device Design
Action
Ziehm Imaging, Inc planned action to bring these defects into compliance: 1. You will contact customers and initiate a software update to correct the defects. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call (866) 949-4346.
Quantity in commerce
1 device
Distribution
US Distribution including Puerto Rico and to the states of :FL, GA, VA,MA. ME, MI, NY, NJ, PA, RI, IL, TX, AZ, CA, NV, UT, and WA.
Product codes and lots
Serial # 10841.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2642-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2639-2016Class IITerminatedVoluntary: Firm initiated
Z-2640-2016Class IITerminatedVoluntary: Firm initiated
Z-2641-2016Class IITerminatedVoluntary: Firm initiated
Z-2642-2016Class IITerminatedVoluntary: Firm initiated
Z-2643-2016Class IITerminatedVoluntary: Firm initiated
Z-2644-2016Class IITerminatedVoluntary: Firm initiated