Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device...
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) began a recall of Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument on . FDA classified it as Class II and gives the reason as "Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1829-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
- Product
- Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument.
- Product code
- KWP Appliance, Fixation, Spinal Interlaminal
- Reason for recall
- Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- RTI sent an Urgent Voluntary Product Notification dated April 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. RTI Surgical is conducting a voluntary recall of the Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve due to the potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip resulting in possible patient injury. For further questions, please call (906) 226-4489.
- Quantity in commerce
- 17 devices
- Distribution
- US nationwide distribution in CA, FL, IN, MI, MO, NC, NY, and TX.
- Product codes and lots
- Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve; Lot Number: 170684; Part Number: 26-COUNTER-T-XLINK
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1829-2016 | Class II | Terminated | Voluntary: Firm initiated |