Z-1829-2016Class II

Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device...

RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) / initiated 2016-04-22 / Terminated / root cause: manufacturing process

RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) began a recall of Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument on . FDA classified it as Class II and gives the reason as "Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1829-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
Product
Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument.
Product code
KWP Appliance, Fixation, Spinal Interlaminal
Reason for recall
Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
RTI sent an Urgent Voluntary Product Notification dated April 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. RTI Surgical is conducting a voluntary recall of the Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve due to the potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip resulting in possible patient injury. For further questions, please call (906) 226-4489.
Quantity in commerce
17 devices
Distribution
US nationwide distribution in CA, FL, IN, MI, MO, NC, NY, and TX.
Product codes and lots
Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve; Lot Number: 170684; Part Number: 26-COUNTER-T-XLINK

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1829-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1829-2016Class IITerminatedVoluntary: Firm initiated