510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1814-2016Class II

PerFuse Decompression Instrument PN: 800-0541, Lot:100650. Designed to access the femoral head for...

Zimmer GmbH / initiated 2016-04-11 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of PerFuse Decompression Instrument PN: 800-0541, Lot:100650. Designed to access the femoral head for decompression on . FDA classified it as Class II and gives the reason as "The Trocar and Plunger Assemblies are missing from the instrument, which could result in a delay in surgery greater than 30 minutes". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1814-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
PerFuse Decompression Instrument PN: 800-0541, Lot:100650. Designed to access the femoral head for decompression.
Product code
FMF Syringe, Piston
Reason for recall
The Trocar and Plunger Assemblies are missing from the instrument, which could result in a delay in surgery greater than 30 minutes.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Zimmer Biomet has initiated a voluntary recall of the PerFuse Decompression Instrument, following an investigation which identified that the Trocar and Plunger Assemblies are missing from the instrument. The firm contacted customers via telephone and e-mail on March 28, 2016; and official recall notices were mailed via FedEx on 4/11/2016. Customers were instructed to quarantine recalled items, Complete the Certificate of Acknowledgment and email a copy to [email protected]; Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA: www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or Call (800)FDA-1088 . Questions related to this recall should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Quantity in commerce
22
Distribution
Distributed in the US states of CA, GA, MN, and OH, and in the Netherlands.
Product codes and lots
PerFuse Decompression Instrument PN: 800-0541, Lot:100650

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1814-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1814-2016Class IITerminatedVoluntary: Firm initiated