Z-1814-2016Class II
PerFuse Decompression Instrument PN: 800-0541, Lot:100650. Designed to access the femoral head for...
Zimmer GmbH / initiated 2016-04-11 / Terminated / root cause: manufacturing process
Biomet, Inc. began a recall of PerFuse Decompression Instrument PN: 800-0541, Lot:100650. Designed to access the femoral head for decompression on . FDA classified it as Class II and gives the reason as "The Trocar and Plunger Assemblies are missing from the instrument, which could result in a delay in surgery greater than 30 minutes". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1814-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- PerFuse Decompression Instrument PN: 800-0541, Lot:100650. Designed to access the femoral head for decompression.
- Product code
- FMF Syringe, Piston
- Reason for recall
- The Trocar and Plunger Assemblies are missing from the instrument, which could result in a delay in surgery greater than 30 minutes.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Zimmer Biomet has initiated a voluntary recall of the PerFuse Decompression Instrument, following an investigation which identified that the Trocar and Plunger Assemblies are missing from the instrument. The firm contacted customers via telephone and e-mail on March 28, 2016; and official recall notices were mailed via FedEx on 4/11/2016. Customers were instructed to quarantine recalled items, Complete the Certificate of Acknowledgment and email a copy to [email protected]; Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA: www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or Call (800)FDA-1088 . Questions related to this recall should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
- Quantity in commerce
- 22
- Distribution
- Distributed in the US states of CA, GA, MN, and OH, and in the Netherlands.
- Product codes and lots
- PerFuse Decompression Instrument PN: 800-0541, Lot:100650
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1814-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY