Various trauma and sports medicine instruments and implants. Screw, fixation, bone
Biomet, Inc. began a recall of Various trauma and sports medicine instruments and implants. Screw, fixation, bone on . FDA classified it as Class II and gives the reason as "Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1823-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Various trauma and sports medicine instruments and implants. Screw, fixation, bone.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 1, 2016, to all affected customers. Customers were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
- Distribution
- Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
- Product codes and lots
- Part #'s HAA60-11030 HAA60-13030 HAA60-15040 HAA60-16040 HAA60-18040 Lot #'s 016150 016210 165130 016230 133890 181600 133970 134000 181620
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1815-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1816-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1817-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1818-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1819-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1820-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1821-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1822-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1823-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1824-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1825-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1826-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1827-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1828-2016 | Class II | Terminated | Voluntary: Firm initiated |