Z-0005-2017Class I
HeartWare Ventricular Assist System Controller
HeartWare, Inc. / initiated 2016-06-08 / Terminated / root cause: design
Heartware, Inc. began a recall of HeartWare Ventricular Assist System Controller Product Usage: For use as a bridge-to-cardiac transplantation in patients who are at risk of death from... on . FDA classified it as Class I and gives the reason as "Loose connector ports in controller body". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0005-2017
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- HeartWare, Inc.
- Product
- HeartWare Ventricular Assist System Controller Product Usage: For use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HVAD is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- Loose connector ports in controller body.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- HeartWare sent an Urgent Medical Device Correction letter dated June 8, 2016, to all affected customers, HeartWare advised clinicians to remind patients to follow unit instructions and water avoidance, and to regularly schedule appointments to check the integrity of the unit and to be cautious with connectors when inspecting connectors. Clinicians were also instructed to send back the enclosed "Acknowledgement Form" within 30 days and to forward the message to necessary individuals within their organization. Customers with questions were instructed to contact their local HeartWare representative. For questions regarding this recall call 305-364-1575.
- Quantity in commerce
- 4235 Worldwide, 7272 units in U.S.
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI and countries of Australia, Austria, Belgium, Canada, France, Germany, Lebanon, Malaysia, Sweden, Switzerland, & United Kingdom.
- Product codes and lots
- All Heartware Controllers manufactured after September 2014. US Model No: 1400, 1401XX, 1403US. International Model No: 1400, 1401XX, 1407XX.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0005-2017 | Class I | Terminated | Voluntary: Firm initiated |