510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1926-2016Class II

Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable...

Medtronic Neuromodulation / initiated 2016-05-03 / Terminated

Medtronic Neuromodulation began a recall of Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of... on . FDA classified it as Class II and gives the reason as "Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1926-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.
FDA root cause
Under Investigation by firm
Action
Consignees were sent on 5/3/2016 a Medtronic "Urgent Medical Device Safety notification" letter dated May 2016. The letter was addressed to "Healthcare Professional". The letter described the problem and the product involved in the recall. The letter listed the recommendation to patients to follow current recharge instructions, paying particular attention to Charging Efficiency and Battery Charge Level indicators on the recharger. Consignees were also requested to complete and return the "Physician Reply Form:. For questions contact Medtronic Neuromodulation Technical services at 1-800-707-0933 weekdays 7 am - 6 pm CST.
Quantity in commerce
60,941
Distribution
Worldwide Distribution
Product codes and lots
All lot/serial numbers are affected for both Model numbers 37714 and 97714.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1926-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1926-2016Class IITerminatedVoluntary: Firm initiated