Stryker FlowPort II Adapter: Model number: 00CAT00778 The FlowPort II Adapter is intended to connect the...
Stryker Corporation began a recall of Stryker FlowPort II Adapter: Model number: 00CAT00778 The FlowPort II Adapter is intended to connect the FlowPort II Cannulas to commercially available, 4mm... on . FDA classified it as Class II and gives the reason as "Complaints were received for the Stryker FlowPort II Adapter, and investigation found that the scope sat in the adapter 180 degrees in the wrong direction". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1910-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker FlowPort II Adapter: Model number: 00CAT00778 The FlowPort II Adapter is intended to connect the FlowPort II Cannulas to commercially available, 4mm arthroscopes in surgical procedures.
- Product code
- NBH Accessories, Arthroscopic
- Reason for recall
- Complaints were received for the Stryker FlowPort II Adapter, and investigation found that the scope sat in the adapter 180 degrees in the wrong direction
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A recall letter was not sent as all affected devices have been returned. A ship hold was placed on the device on October 16, 2015.
- Quantity in commerce
- 21 units
- Distribution
- GA, UT, MD, NJ, MT
- Product codes and lots
- Lot Numbers Affected: 13604
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1910-2016 | Class II | Terminated | Voluntary: Firm initiated |