510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1912-2016Class II

DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43

Synthes (USA) Products LLC / initiated 2016-05-02 / Terminated / root cause: manufacturing process

Synthes (USA) Products LLC began a recall of DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43 on . FDA classified it as Class II and gives the reason as "Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1912-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43
Product code
HTO Reamer
Reason for recall
Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points. This could potentially be reproduced during use and reprocessing.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Depuy Synthes sent an Urgent Notice: Medical Device Recall letter on 05/02/2016 to all affected consignees. Consignees were instructed to do the following: review your inventory immediately to see if any of the affected products were being stored, and call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number and replacement order, to complete the Verification Section (page 3) and return with the product to Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132.
Quantity in commerce
244 units
Distribution
US nationwide distribution.
Product codes and lots
Catalog ID # 351.43 Lot numbers: XXX5071, 3945479, 4738380, XXX5072, 3997000, 4816738, XXX5073, 3996999, 4816739, A3CN449, 3997338, 4441056, 3007721, 4326050, 5407442, 3148892, 4307973, 5582819, 3175916, 4387338, 5722664, 3175202, 4441055, 5813837, 3175201, 4441079, 5786778, 3921562, 4559672, 5800844, 3928332, 4561383, 5903413, 3935789, 4619549, 5910995.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1912-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1912-2016Class IITerminatedVoluntary: Firm initiated