DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43
Synthes (USA) Products LLC began a recall of DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43 on . FDA classified it as Class II and gives the reason as "Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1912-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes (USA) Products LLC
- Product
- DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument. Part number 351.43
- Product code
- HTO Reamer
- Reason for recall
- Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points. This could potentially be reproduced during use and reprocessing.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Depuy Synthes sent an Urgent Notice: Medical Device Recall letter on 05/02/2016 to all affected consignees. Consignees were instructed to do the following: review your inventory immediately to see if any of the affected products were being stored, and call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number and replacement order, to complete the Verification Section (page 3) and return with the product to Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132.
- Quantity in commerce
- 244 units
- Distribution
- US nationwide distribution.
- Product codes and lots
- Catalog ID # 351.43 Lot numbers: XXX5071, 3945479, 4738380, XXX5072, 3997000, 4816738, XXX5073, 3996999, 4816739, A3CN449, 3997338, 4441056, 3007721, 4326050, 5407442, 3148892, 4307973, 5582819, 3175916, 4387338, 5722664, 3175202, 4441055, 5813837, 3175201, 4441079, 5786778, 3921562, 4559672, 5800844, 3928332, 4561383, 5903413, 3935789, 4619549, 5910995.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1912-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.