Z-1695-2016Class II
Philips Healthcare DigitalDiagnost stationary X-ray system
Philips North America LLC / initiated 2015-11-10 / Terminated
Philips Healthcare began a recall of Philips Healthcare DigitalDiagnost stationary X-ray system on . FDA classified it as Class II and gives the reason as "The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1695-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Healthcare DigitalDiagnost stationary X-ray system
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Philips Healthcare plans to correct the defect at no cost to customers by implementing a field correction involving the installation of new software. 1. You will contact customers and initiate a software update to correct the defect. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call 1-800- 722-9377.
- Quantity in commerce
- 160
- Distribution
- USA (nationwide) Distribution to the states of : ( DuraDiagnost) to the states of : OH, CA, WA and NE. (DigitalDiagnost) to the states of : AK, AZ, AL, CA, CT, CO, DE, FL, GA, HI, ID, KY, LA, MA, MN, MD, MS, ME, MO, NE, NY, ND, OH and PA.
- Product codes and lots
- software version - 4.0.4, 4.1.2, 4.1.3.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1695-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1696-2016 | Class II | Terminated | Voluntary: Firm initiated |