510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1696-2016Class II

Philips Healthcare DuraDiagnost stationary X-ray system

Philips North America LLC / initiated 2015-11-10 / Terminated

Philips Healthcare began a recall of Philips Healthcare DuraDiagnost stationary X-ray system on . FDA classified it as Class II and gives the reason as "The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1696-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Healthcare DuraDiagnost stationary X-ray system
Product code
KPR System, X-Ray, Stationary
Reason for recall
The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image.
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips Healthcare plans to correct the defect at no cost to customers by implementing a field correction involving the installation of new software. 1. You will contact customers and initiate a software update to correct the defect. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call 1-800- 722-9377.
Quantity in commerce
5
Distribution
USA (nationwide) Distribution to the states of : ( DuraDiagnost) to the states of : OH, CA, WA and NE. (DigitalDiagnost) to the states of : AK, AZ, AL, CA, CT, CO, DE, FL, GA, HI, ID, KY, LA, MA, MN, MD, MS, ME, MO, NE, NY, ND, OH and PA.
Product codes and lots
software version - 4.0.2, 4.0.3, 4.0.4 into 4.0.5

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1696-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1695-2016Class IITerminatedVoluntary: Firm initiated
Z-1696-2016Class IITerminatedVoluntary: Firm initiated