Z-1896-2016Class I

Gas Delivery Engine (GDE), 16650A A continuous ventilator intended to mechanically control or assist broad...

Jaeger Medical GmbH / initiated 2016-05-17 / Terminated

Carefusion 211 Inc dba Carefusion began a recall of Gas Delivery Engine (GDE), 16650A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a... on . FDA classified it as Class I and gives the reason as "CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1896-2016
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Jaeger Medical GmbH
Product
Gas Delivery Engine (GDE), 16650A A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
FDA root cause
Under Investigation by firm
Action
CareFusion sent a Field Safety Notice dated May 17, 2016, to all affected customers. The notice informed customers that CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM). This malfunction creates a blank AVEA Ventilator UIM and INOP condition during pre-use check or during use with cessation of ventilation to the patient. By design an audible clinical alarm is activated. Customers were informed of the problems identified and the actions to be taken. Customers were not required to return the affected devices or suspend use, customers would receive an copy of the identified affected serial numbers, as well as a response card, customers were requested to promptly return the response card to expedite the correction process and acknowledge receipt of the notification, Customers would be contacted by CareFusion to arrange onsite remediation of the affected devices, in the interim if any AVEA ventilator unit in their facility exhibits a blank AVEA Ventilator UIM and INOP condition during pre-use check or during use with cessation of ventilation, immediately remove the ventilator from service, provide alternate ventilation and contact CareFusion Technical Support per the contact information listed below to report the issue. Customers with questions should call 888-562-6016, for technical support 800-231-2466 or [email protected].
Quantity in commerce
922 units total (501 units in US)
Distribution
Worldwide Distribution - US (Nationwide) and the countries of Australia, Hungary, Paraguay, Belgium, Indonesia, Philippines, Bolivia, Ireland, Poland, Bermuda, India, Saudi Arabia, Canada, Japan, Singapore, Chile, Jordon, South Africa, China, Kuwait, Taiwan, Ecuador, Malaysia, Thailand, Egypt, Mexico, Turkey, France, Netherlands, United Kingdom, Guatemala, and Nicaragua.
Product codes and lots
n/a

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1896-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1888-2016Class ITerminatedVoluntary: Firm initiated
Z-1889-2016Class ITerminatedVoluntary: Firm initiated
Z-1890-2016Class ITerminatedVoluntary: Firm initiated
Z-1891-2016Class ITerminatedVoluntary: Firm initiated
Z-1892-2016Class ITerminatedVoluntary: Firm initiated
Z-1893-2016Class ITerminatedVoluntary: Firm initiated
Z-1894-2016Class ITerminatedVoluntary: Firm initiated
Z-1895-2016Class ITerminatedVoluntary: Firm initiated
Z-1896-2016Class ITerminatedVoluntary: Firm initiated
Z-1897-2016Class ITerminatedVoluntary: Firm initiated
Z-1898-2016Class ITerminatedVoluntary: Firm initiated
Z-1899-2016Class ITerminatedVoluntary: Firm initiated