Z-1965-2016Class II

TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is...

Integra LifeSciences Corporation / initiated 2016-05-10 / Terminated / root cause: manufacturing process

Integra LifeSciences Corp. began a recall of TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw... on . FDA classified it as Class II and gives the reason as "As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1965-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.
Product code
HRS Plate, Fixation, Bone
Reason for recall
As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The firm, Integra Lifesciences, sent an "Urgent: Voluntary Medical Device Recall" letter dated 5/10/16 to affected consignees/customers via traceable courier service on 5/10/16. The letter describes the product, problem and actions to be taken. The customers were instructed to stop using and distributing product immediately; complete and return the Recall Acknowledgement and Return Form via fax to 1-609-275-9445 or email to [email protected], even if you do not have any of the product. Should you have any questions, please contact Customer Service at 1-866-854-8300.
Quantity in commerce
71 units
Distribution
US Distribution to states of: CO, FL, IL, MI, NJ, NY, SC, TX, and VA.
Product codes and lots
Lot UP15H2, Lot UP09D3, Lot UP07E2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1965-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1965-2016Class IITerminatedVoluntary: Firm initiated