Persona Trabecular Metal Tibia; Persona Trabecular Metal Tibial Plate Instruments and Modular Brackets; Used...
Zimmer Biomet, Inc. began a recall of Persona Trabecular Metal Tibia; Persona Trabecular Metal Tibial Plate Instruments and Modular Brackets; Used to facilitate the implantation of joint... on . FDA classified it as Class II and gives the reason as "Zimmer Inc. is initiating a recall of specific Persona" Trabecular Metal" Tibial Plate Instruments and Modular Brackets, due to incidents of radiolucent lines and loosening in implants". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1917-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Persona Trabecular Metal Tibia; Persona Trabecular Metal Tibial Plate Instruments and Modular Brackets; Used to facilitate the implantation of joint replacement implants.
- Product code
- MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Reason for recall
- Zimmer Inc. is initiating a recall of specific Persona" Trabecular Metal" Tibial Plate Instruments and Modular Brackets, due to incidents of radiolucent lines and loosening in implants.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Zimmer Inc. is initiating a voluntary recall of Persona" Trabecular Metal" Tibia / Tibial Plate Instruments and Modular Brackets following an increase in complaints of radiolucent lines and loosening, and earlier recall of the Persona" Trabecular Metal" Tibial Plates . All sizes and lots of the affected devices are being removed from distribution. Recall notices were issued to affected distributors, hospitals, and surgeons on 05/11/2015 via mail. Customers may contact the following with any questions: 1-877-946- 2761 between 8:00 am and 5:00pm EST. Customers are asked to do the following: 1. Review the notification and ensure affected personnel are aware of the contents. 2. Assist Zimmer Biomet sales representative with the quarantine of any affected product. 3. Zimmer Biomet sales representative will remove the recalled product from customer facilities. 4. Complete the Certificate of Acknowledgement Form and return to: [email protected].
- Quantity in commerce
- 10,256
- Distribution
- Worldwide Distribution -- US, including the states of AK, AL, AZ, CA, FL, GA, HI, ID, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, and WI; Foreign: Australia, New Zealand, Korea, Austria, Belgium, Germany, Italy, France, Luxembourg, Netherlands, South Africa, Spain, Sweden, Switzerland, and United Arab Emirates.
- Product codes and lots
- Part Number, Lot Number, Date of Manufacture: 42509909210, 63138589, 07/28/2015; 42509909210, 63097616, 06/10/2015; 00590708181, 56595825, 04/20/2015; 00590708185, 56610423, 03/12/2015; 00590708176, 56595826, 03/04/2015; 00590708179, 56595827, 03/04/2015; 00590708180, 56595828, 03/04/2015; 00590708181, 56595824, 03/02/2015; 00590708178, 56595692, 02/17/2015; 00590708176, 56595691, 02/16/2015; 00590708177, 56595801, 02/16/2015; 00590708180, 56595694, 02/04/2015; 00590708179, 56595693, 01/29/2015; 00590708181, 56595642, 01/20/2015; 00590708177, 56595625, 01/06/2015; 42539807902, 62906456, 12/14/2014; 00590708178, 56595569, 12/12/2014; 42539808802, 62885379, 11/22/2014; 42539807102, 62885377, 11/21/2014; 42539806702, 62885376, 11/20/2014; 42539806701, 62885375, 11/18/2014; 42539808801, 62863647, 10/29/2014; 42539802000, 62854350, 10/16/2014; 42539808802, 62849167, 10/14/2014; 42539806702, 62849168, 10/13/2014; 42539807502, 62839311, 10/08/2014; 42539807101, 62833614, 10/07/2014; 42539802000, 62844358, 10/07/2014; 42539808301, 62839310, 10/04/2014; 42539806401, 62833616, 10/03/2014; 42539807101, 62839312, 10/03/2014; 42539800500, 62844377, 09/30/2014; 42539802000, 62827692, 09/29/2014; 42539807901, 62831188, 09/25/2014; 42539808301, 62833610, 09/25/2014; 00590708176, 56595440, 09/25/2014; 42539806401, 62837612, 09/24/2014; 42539807502, 62833612, 09/24/2014; 00590708179, 56595441, 09/24/2014; 00590708180, 56595442, 09/24/2014; 42539807501, 62831191, 09/22/2014; 00590708181, 56595443, 09/18/2014; 42539807501, 62827408, 09/15/2014; 42539807901, 62827407, 09/15/2014; 42509909210, 62817496, 09/08/2014; 42539806402, 62817469, 09/03/2014; 42539808302, 62817467, 09/03/2014; 42539807902, 62817468, 09/02/2014; 42539806701, 62813822, 08/24/2014; 00590708177, 56595435, 08/05/2014; 42539802000, 62769875, 08/03/2014; 42539800500, 62779559, 07/30/2014; 42539800100, 62779557, 07/29/2014; 42539800700, 62769874, 07/28/2014; 42539800100, 62769871, 07/27/2014; 42539800500, 62769872, 07/27/2014; 42509909210, 62782990, 07/24/2014; 00590708176, 56595308, 07/17/2014; 42539807101, 62741938, 07/07/2014; 42539808802, 62736082, 06/26/2014; 42539807501, 62736060, 06/25/2014; 42539808801, 62736081, 06/25/2014; 42539807502, 62732180, 06/19/2014; 42539806401, 62724757, 06/14/2014; 42539806702, 62736062, 06/14/2014; 42539807501, 62696888, 05/14/2014; 42539807901, 62702760, 05/05/2014; 42539808301, 62702762, 05/02/2014; 42539807102, 62665087, 04/09/2014; 42509909210, 62658516, 04/03/2014; 42539807902, 62677171, 04/02/2014; 42539806701, 62677173, 04/01/2014; 42539806402, 62677172, 04/01/2014; 42539808302, 62677174, 04/01/2014; 42539800100, 62550569, 12/14/2013; 42539800700, 62550582, 12/13/2013; 42509909210, 62563823, 12/09/2013; 42539800500, 62550573, 11/23/2013; 42539800100, 62550565, 11/18/2013; 42539800700, 62550578, 11/18/2013; 42539802000, 62550588, 11/13/2013; 42539800500, 62519720, 11/11/2013; 42539802000, 62485450, 09/03/2013; 42539807101, 62474901, 08/22/2013; 42539808801, 62469464, 08/20/2013; 42539808802, 62469457, 08/20/2013; 42509909210, 62468258, 08/20/2013; 00590708186, 56524701, 08/20/2013; 42539807501, 62469447, 08/16/2013; 00590708186, 56524691, 08/16/2013; 42539807901, 62469448, 08/15/2013; 42539807502, 62469449, 08/15/2013; 00590708185, 56529311, 08/14/2013; 42509909210, 62460135, 08/02/2013; 42539806702, 62449207, 08/01/2013; 42539806401, 62424809, 07/26/2013; 42539807502, 62424810, 07/25/2013; 42539808802, 62429189, 07/25/2013; 42539807501, 62429187, 07/24/2013; 00590708185, 56570277, 07/17/2013; 42509909210, 62442315, 07/16/2013; 00590708186, 56524662, 07/16/2013; 00590708185, 56570276, 07/12/2013; 42539807102, 62420235, 07/08/2013; 42539800100, 62434639, 07/08/2013; 42539800500, 62434641, 07/08/2013; 42539800700, 62434642, 07/08/2013; 42509909210, 62429474, 06/28/2013; 42539807901, 62419053, 06/27/2013; 42539807101, 62420230, 06/26/2013; 42539807101, 62391499, 06/25/2013; 42539807901, 62391505, 06/25/2013; 00590708186, 56524634, 06/21/2013; 42539807902, 62386403, 06/20/2013; 42539808801, 62377922, 06/17/2013; 42539806402, 62377911, 06/17/2013; 42539808301, 62377914, 06/12/2013; 42539808302, 62377918, 06/12/2013; 42539806401, 62385935, 06/11/2013; 42539806701, 62386402, 06/11/2013; 00590708179, 56528980, 06/11/2013; 42539802000, 62364964, 06/07/2013; 42509909210, 62412194, 06/07/2013; 42539800700, 62357622, 06/03/2013; 42539807101, 62357870, 05/24/2013; 42539807901, 62366122, 05/24/2013; 42539806702, 62366119, 05/24/2013; 42509909210, 62399872, 05/24/2013; 42539800500, 62315770, 05/18/2013; 42539800100, 62386789, 05/16/2013; 42539800500, 62315769, 05/16/2013; 42539800100, 62301872, 05/15/2013; 00590708178, 56528979, 05/11/2013; 42539807902, 62366123, 05/09/2013; 42509909210, 62383744, 05/07/2013; 42539807501, 62301873, 05/06/2013; 42539807102, 62315772, 05/06/2013; 42539807502, 62301874, 05/06/2013; 42539807902, 62301875, 05/06/2013; 00590708176, 56528977, 05/02/2013; 00590708180, 56528981, 05/01/2013; 00590708181, 56528982, 05/01/2013; 00590708177, 56528978, 04/30/2013; 42539808802, 62374294, 04/25/2013; 42509909210, 62369694, 04/22/2013; 42539808802, 62357874, 04/17/2013; 42539808802, 62357873, 04/17/2013; 42539808801, 62357872, 04/16/2013; 42539808801, 62357871, 04/15/2013; 42509909210, 62357131, 04/04/2013; 42539808302, 62290686, 03/29/2013; 42509909210, 62345129, 03/22/2013; 00590708185, 56529104, 03/22/2013; 00590708179, 56519351, 03/20/2013; 00590708186, 56524431, 03/12/2013; 00590708186, 56524539, 03/11/2013; 00590708178, 56519339, 03/09/2013; 42509909210, 62331503, 03/07/2013; 00590708186, 56524430, 03/05/2013; 00590708177, 56519352, 02/26/2013; 00590708176, 56519335, 02/19/2013; 00590708178, 56519338, 02/19/2013; 00590708180, 56519348, 02/18/2013; 42539807501, 62274305, 02/14/2013; 42539806402, 62276254, 02/14/2013; 00590708179, 56519350, 02/13/2013; 00590708181, 56519337, 02/13/2013; 42539807502, 62274296, 02/05/2013; 42539807902, 62274276, 02/05/2013; 42539802000, 62274308, 02/05/2013; 42539806701, 62292435, 01/29/2013; 42539808301, 62289365, 01/24/2013; 42539807901, 62281193, 01/21/2013; 42539807101, 62256180, 01/15/2013; 42539806702, 62256179, 01/15/2013; 42539808302, 62276680, 01/15/2013; 42539800100, 62264917, 01/15/2013; 42539800500, 62264918, 01/15/2013; 42539800700, 62264919, 01/15/2013; 42539807102, 62254389, 01/14/2013; 42539800700, 62264920, 01/14/2013; 42539806402, 62248613, 12/05/2012; 42539806401, 62248612, 12/03/2012; 42539806701, 62248614, 12/03/2012; 00590708179, 56519203, 12/03/2012; 42539800100, 62244002, 11/26/2012; 42539800500, 62244003, 11/26/2012; 42539800700, 62244004, 11/26/2012; 00590708176, 56519244, 11/26/2012; 00590708178, 56519210, 11/26/2012; 00590708180, 56519219, 11/26/2012; 00590708177, 56519247, 11/24/2012; 00590708179, 56519204, 11/24/2012; 00590708180, 56519217, 11/24/2012; 00590708180, 56519218, 11/24/2012; 00590708181, 56519207, 11/24/2012; 00590708178, 56519209, 11/20/2012; 42539806701, 62237938, 11/19/2012; 42539808301, 62239035, 11/19/2012; 42539806402, 62237937, 11/19/2012; 42539808302, 62239029, 11/19/2012; 42539807101, 62233958, 11/16/2012; 42539807902, 62237932, 11/16/2012; 00590708176, 56519243, 11/16/2012; 00590708181, 56519206, 11/16/2012; 42539807501, 62237929, 11/15/2012; 42539807901, 62233959, 11/15/2012; 42539808301, 62237935, 11/15/2012; 42539807102, 62237962, 11/15/2012; 42539807502, 62237886, 11/15/2012; 42539808302, 62237936, 11/15/2012; 42539802000, 62233960, 11/15/2012; 00590708177, 56519246, 11/15/2012; 42539806401, 62226724, 11/14/2012; 42539806701, 62224797, 11/14/2012; 42539806402, 62224795, 11/14/2012; 00590708176, 56519242, 11/07/2012; 00590708178, 56519208, 11/07/2012; 00590708179, 56519202, 11/07/2012; 00590708181, 56519205, 11/07/2012; 00590708177, 56519245, 11/06/2012; 42539806401, 62214944, 10/22/2012; 42539806702, 62214945, 10/18/2012;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1917-2016 | Class II | Terminated | Voluntary: Firm initiated |