Z-1716-2016Class II

Philips Healthcare Ingenuity Core Computed Tomography X-Ray System

Philips North America LLC / initiated 2015-09-06 / Terminated / root cause: software

Philips Healthcare began a recall of Philips Healthcare Ingenuity Core Computed Tomography X-Ray System on . FDA classified it as Class II and gives the reason as "Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1716-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Healthcare Ingenuity Core Computed Tomography X-Ray System
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips plans to initiate a field corrective action that will install a new software release to address the issue. Philips will modify the systems in the field to correct the defect at no cost to the customer. 1. You will contact customers and initiate a software update to correct the defect. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call (1-800) 722-9377, Option 5
Quantity in commerce
US-423
Distribution
US Distribution - AZ, PA, FL, DC, W VA, NC, SC, AR, OK, OH, NY, NV, TX, PA, IN, CA, MN, MA, DC, MS, MD, OR, IL, CO, MO, VT, IA, AL, LA, GA and WA.
Product codes and lots
Software v. 4.13, 4.14 and 4.15.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1716-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1713-2016Class IITerminatedVoluntary: Firm initiated
Z-1714-2016Class IITerminatedVoluntary: Firm initiated
Z-1715-2016Class IITerminatedVoluntary: Firm initiated
Z-1716-2016Class IITerminatedVoluntary: Firm initiated
Z-1717-2016Class IITerminatedVoluntary: Firm initiated
Z-1718-2016Class IITerminatedVoluntary: Firm initiated
Z-1719-2016Class IITerminatedVoluntary: Firm initiated