Z-2082-2016Class II

Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028

Medtronic Navigation, Inc / initiated 2016-03-17 / Terminated

Medtronic Navigation, Inc.-Littleton began a recall of Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028 on . FDA classified it as Class II and gives the reason as "It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2082-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028
Product code
OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
Reason for recall
It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.
FDA root cause
Radiation Control for Health and Safety Act
Action
Medtronic's planned action to bring device into Compliance. 1. You will provide an Errata sheet to customers updating the missing and incorrect information. 2. Field Service Engineers (FSEs) will perform radiation output measurements via validated procedures to attain compliance with 21CFR part 1020.32(k)(6). 3. You will inspect and test X-ray generator and system motion batteries for damage, in accordance with the plan submitted to CDRH. Damaged batteries will be replaced. 4. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet. For further questions, please call (800) 595-9709.
Quantity in commerce
US - 581
Distribution
Nationwide Distribution
Product codes and lots
Model No. BI-700-00027 and BT-700-00028

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2082-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2082-2016Class IITerminatedVoluntary: Firm initiated