Z-2082-2016Class II
Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028
Medtronic Navigation, Inc / initiated 2016-03-17 / Terminated
Medtronic Navigation, Inc.-Littleton began a recall of Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028 on . FDA classified it as Class II and gives the reason as "It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2082-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Medtronic's planned action to bring device into Compliance. 1. You will provide an Errata sheet to customers updating the missing and incorrect information. 2. Field Service Engineers (FSEs) will perform radiation output measurements via validated procedures to attain compliance with 21CFR part 1020.32(k)(6). 3. You will inspect and test X-ray generator and system motion batteries for damage, in accordance with the plan submitted to CDRH. Damaged batteries will be replaced. 4. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet. For further questions, please call (800) 595-9709.
- Quantity in commerce
- US - 581
- Distribution
- Nationwide Distribution
- Product codes and lots
- Model No. BI-700-00027 and BT-700-00028
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2082-2016 | Class II | Terminated | Voluntary: Firm initiated |