Medtronic Navigation O-arm 02 surgical x-ray imaging system
Medtronic Navigation, Inc.-Littleton began a recall of Medtronic Navigation O-arm 02 surgical x-ray imaging system on . FDA classified it as Class II and gives the reason as "Review of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2083-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic Navigation O-arm 02 surgical x-ray imaging system
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Review of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR part 1 020.30(h).
- Root cause tag
- labeling and UDI
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Medtronic planned action to bring device into Compliance: 1. You will contact customers and provide an Errata sheet to them updating the missing and incorrect information. 2. Medtronic field service personnel will perform radiation output measurements via validated procedures to attain compliance with 21CRF part 1020.32(k)(6). 3. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet. For questions call (800)-595-9709.
- Quantity in commerce
- US - 63
- Distribution
- Nationwide Distribution
- Product codes and lots
- model number BT-700-02000
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2083-2016 | Class II | Terminated | Voluntary: Firm initiated |