Z-2083-2016Class II

Medtronic Navigation O-arm 02 surgical x-ray imaging system

Medtronic Navigation, Inc / initiated 2016-03-17 / Terminated / root cause: labeling and UDI

Medtronic Navigation, Inc.-Littleton began a recall of Medtronic Navigation O-arm 02 surgical x-ray imaging system on . FDA classified it as Class II and gives the reason as "Review of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2083-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Navigation O-arm 02 surgical x-ray imaging system
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Review of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR part 1 020.30(h).
Root cause tag
labeling and UDI
FDA root cause
Radiation Control for Health and Safety Act
Action
Medtronic planned action to bring device into Compliance: 1. You will contact customers and provide an Errata sheet to them updating the missing and incorrect information. 2. Medtronic field service personnel will perform radiation output measurements via validated procedures to attain compliance with 21CRF part 1020.32(k)(6). 3. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet. For questions call (800)-595-9709.
Quantity in commerce
US - 63
Distribution
Nationwide Distribution
Product codes and lots
model number BT-700-02000

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2083-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2083-2016Class IITerminatedVoluntary: Firm initiated